Buffer 2 - Hach POLYMETRON 8362 User Manual

Ph measurement
Hide thumbs Also See for POLYMETRON 8362:
Table of Contents

Advertisement

Transmitter 9135 - pH measurement in pure water
pH CALIB.

BUFFER 2

Press
ENTER
pH CALIB.
6.98
pH
0
06 . 98 pH
Page 38
Buffer 2
Maintain the screw nut in position and unscrew the
chamber (A).
Empty out the water contained in the chamber,
and rinse with demineralised water.
Refill with a fresh buffer (B) pH 9.00, re-screw the
nut (C).
Validate by Enter.
The temperature indication disappears and the pH
measure appears.
The measured temperature by means of the 8362
temperature probe system allows to adjust the
value of the reference solution.
Utilize the
Enter. Press directly Enter to not change a digit.
After the last number, press Enter to validate the
calibration.
Calibration parameters are displayed.
Put the calibration date.
Press on
Utilize
Enter.
Press Enter after validation of the last digit of the
year. Press another time Enter to validate the
calibration parameters.
Note :
If you quit with out validate the calibration
parameters, the transmitter will set the last
calibration and not yours.