Transmitter 9135 - pH measurement in pure water

PARAMETERS Menu

PARAMETERS
DATE
: 26/03/96
SLOPE
: 95%
ZERO
: 0,3 pH
ΔT
: 0,2°C

HISTORIC Menu

HISTORIC
DATE
: 26/03/96
SLOPE
: 98%
ZERO
: -0,3 pH
Select.

MAINTENANCE Menu

MAINTENANCE
6.98
pH
26.4 °C
Page 40
The date, slope and zero of the last calibration may
be displayed in the PARAMETERS menu.
The slope and zero parameters of the last two
calibrations may be displayed in the historic menu.
Press the left function key SELECT to display the next
parameters.
When changing an electrode or servicing the
instrument, the transmitter continues to display
measures.
Note :
The
analogue
programmed in the mA menu. The relay state is
not modified.
output
value
is
the
value