S/Measure Menu - Hach POLYMETRON 8362 User Manual

Ph measurement
Hide thumbs Also See for POLYMETRON 8362:
Table of Contents

Advertisement

Transmitter 9135 - pH measurement in pure water
MEASURE
ELECTRODE
TEMP. COMP. :
Select
ELECTRODE
TYPE :
Other
ISO Pt : 1,5 pH
SLOPE : 95 %
Uiso : 200 mV
Page 42

S/MEASURE Menu

ELECTRODE menu allows to set up the electrode
model.
COMP. TEMP menu allows to set up the temperature
compensation.
ELECTRODE
Choice of the electrode.
• Glass
TYPE
:
• Sb
• Redox
• Other
ISO Pt
:
XX,XX pH
The isothermal point corresponds to
the pH value which does not vary
according to the temperature.
SLOPE
:
XX %
Indication of the electrode sensitivity in
% of the theoric value (59,16 mV/pH at
25°C).
Uiso
:
XXXX mV
Uiso is the potential of the isothermal
point.