Transmitter 9135 - pH measurement in pure water
RELAY S1
COEF.XP1 : 100%
T1 : 2.00 s
MIN : 001 i/min
MAX : 200 i/min
Select
RELAY S2
COEF.XP2 : 100%
T2 : 2.00 s
MIN : 001 i/min
MAX : 200 i/min
Select
MANUAL MODE
7.00
25.0 °C
S1
S2
Page 52
S1 & S2 RELAYS
−
Proportionality coefficient : slope of regulation :
COEF.XP1/2 : XXX – In %
−
Value of the pulse width (0,1 < T < 0,7) :
T1/2 :
−
Minimum number of pulses per minute. Program
this value according to the control system
associated to the relay :
MIN :
−
Maximum number of pulses per minute. Program
this value according to the control system
associated to the relay :
MAX :
MANUAL MODE
−
Force relay S1 or S2. Pressing the function key
associated to the relay, the relay is activated till
you release the key. The corresponding icon is
pH
displayed according to the relay state.
PULSE CONTROL :
The pulse period is constant and programmable (3 to
60 s). It is the cycle quotient (α) which varies
according to the sample pH value.
S3
On
Off
XXX – In %
XXX – In i/min
XXX – In i/min
" ?
" ?
T
T
" ?
Time
T