Hach POLYMETRON 8362 User Manual page 47

Ph measurement
Hide thumbs Also See for POLYMETRON 8362:
Table of Contents

Advertisement

Transmitter 9135 - pH measurement in pure water
TEMP. COMP.
MAESURE :
Yes
TYPE : Manual
TEMP. : 20°C
COMP. : Coeff.
Δ. : 00.010 pH
Select
TEMP. COMP.
This menu allows to set up the temperature
compensation.
MEASURE :
- No
- Pt
- Auto.
TYPE
:
- Manual
TEMP.
:
XX °C
COMP.
:
-
NERNST
- Pure
-
Matrix
-
Matrix
-
Matrix
-
Matrix
Coeff.
Δ
XX
:
Choice
of
measurement with or without
Pt100 / Pt1000.
Choice between an automatic or
a manual temperature measu-
rement.
Possibility to enter the sample
temperature
compensation.
Choice
of
the
compensation type :
- linear compensation
(0.1984 mV/°C).
- compensation according to the
ultrapure water curve.
1 - compensation according to the
sulphate
curve
corresponds to a pH 4.0 to
25°C).
2 - compensation according to the
ammoniac/hydrazine
(0.272 mg/l ammoniac + 20 µg/l
corresponds to a pH 9.0 to
25°C).
3 - compensation according to the
ammoniac
/
hydrazine
curve
ammoniac + 10 mg/l morpholine
+ 50 µg/l hydrazine corresponds
to a pH 9.6 to 25°C).
4 - compensation according to the
phosphate
phosphates
ammoniac).
- ajustable coefficient (pH/10°C).
In the case of a programmable
coefficient, possibility to enter
the value of the coefficient
(value in pH/10°C or pH/18°F).
a
temperature
in
a
manual
temperature
(4.84 mg/l
curve
morpholine
/
(1.832 mg/l
curve
(3 mg/l
+
0.3 mg/l
Page 43

Advertisement

Table of Contents
loading

Table of Contents