Hach polymetron 9240 User Manual
Hach polymetron 9240 User Manual

Hach polymetron 9240 User Manual

Multi-channel sodium analyzer
Hide thumbs Also See for polymetron 9240:
Table of Contents

Advertisement

DOC024.52.93034
POLYMETRON Model 9240 Multi-Channel
Sodium Analyzer
USER MANUAL
08/2016, Edition 15

Advertisement

Table of Contents
loading

Summary of Contents for Hach polymetron 9240

  • Page 1 DOC024.52.93034 POLYMETRON Model 9240 Multi-Channel Sodium Analyzer USER MANUAL 08/2016, Edition 15...
  • Page 2: Table Of Contents

    Table of Contents ......................5 Section 1 General Information 1.1 Disclaimer ............................ 5 1.2 Contact information........................5 1.3 Safety information ........................5 1.3.1 Use of hazard information ....................5 1.3.2 Safety recommendations ....................6 1.3.3 Service and repairs ......................6 1.3.4 Potential safety hazards......................
  • Page 3 Table of Contents 4.4.2 RS485 connection ......................45 4.4.3 Input/Output connections ....................47 4.4.4 Sample tubes installation ....................48 4.4.5 Reagents installation ......................49 4.4.6 Magnetic stirrer installation....................50 4.4.7 Reagents volume declaration....................51 4.4.8 Flow rate adjustment ......................52 4.4.9 Sample pH conditioning check ..................53 4.4.10 Reference electrode installation ..................54 4.4.11 Sodium ion selective electrode installation..............56 4.4.12 Fill electrolyte reservoir ....................57 4.5 Analyzer stabilization........................58...
  • Page 4 Table of Contents 7.4.3 Test ........................... 90 7.5 RS485 (or PROFIBUS) ......................90 7.6 Sample channels ........................91 7.6.1 Number of channels ......................91 7.6.2 Channel activation......................92 7.6.3 Sequence .......................... 92 7.6.4 Channel names ......................... 92 ........................93 Section 8 Calibration 8.1 General ............................
  • Page 5 Table of Contents ....................121 Section 12 Default Configuration 12.1 User configuration table ......................121 ..............127 Section 13 Material Safety Data Sheets (MSDS) 13.1 Diisopropylamine (DIPA) ......................127 13.2 Potassium chloride .........................130 13.3 Sodium chloride........................132 13.4 Sodium nitrate ........................134 ....................137 Section 14 Reagent Preparation 14.1 Conditioning Reagent......................137 14.2 Standard Solutions .........................137 14.3 Automatic Calibration Solution (10 ppm Na) ................138...
  • Page 6: Section 1 General Information

    However, Hach Lange assumes no responsibility for any inaccuracies that may be contained in this manual. In no event will Hach Lange be liable for direct, indirect, special, incidental, or consequential damages resulting from any defect or omission in this manual, even if advised of the possibility of such damages.
  • Page 7: Safety Recommendations

    1.3.3 Service and repairs None of the analyzer’s components can be serviced by the user. Only personnel from Hach Lange or its approved representative(s) is (are) authorized to attempt repairs to the system and only components formally approved by the manufacturer should be used.
  • Page 8: Precautionary Labels

    General Information 1.3.5 Precautionary labels Read all labels and tags attached to the analyzer. Personal injury or damage to the analyzer could occur if not observed. This symbol, when noted on a product, indicates a potential hazard which could cause serious personal injury and/or death.
  • Page 9: Product Recycling Information

    General Information 1.4 Product recycling information ENGLISH Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
  • Page 10: General Information

    General Information SVENSKA Elektronikutrustning som är märkt med denna symbol kanske inte kan lämnas in på europeiska offentliga sopstationer efter 2005-08-12. Enligt europeiska lokala och nationella föreskrifter (EU-direktiv 2002/96/EC) måste användare av elektronikutrustning i Europa nu återlämna gammal eller utrangerad utrustning till tillverkaren för kassering utan kostnad för användaren. Obs! Om du ska återlämna utrustning för återvinning ska du kontakta tillverkaren av utrustningen eller återförsäljaren för att få...
  • Page 11: Product Disposal

    European public disposal systems after 12 August 2005. Hach Lange will offer to take back (free of charge to the customer) any old, unserviceable or redundant analyzers and systems which carry the above symbol, and which were originally supplied by Hach Lange.
  • Page 12: Restriction Of Hazardous Substances (Rohs)

    Currently, monitoring and control instruments do not fall within the scope of the RoHS Directive, however Hach Lange has taken the decision to adopt the recommendations in the Directive as the target for all future product design and component purchasing.
  • Page 13 General Information...
  • Page 14: Section 2 Specifications

    Section 2 Specifications 2.1 Technical specifications Specifications are subject to change without notice. Table 1 Technical specifications PERFORMANCE SPECIFICATIONS 0 to 10,000 ppb freely programmable Measuring range 0 to 200 ppm with K-Kit option Non-cationic application: ± 0.1 ppb or ± 5% of reading, whichever is greater Accuracy Cationic application: ±...
  • Page 15: Specifications

    Specifications MECHANICAL SPECIFICATIONS Maximum panel dimension 850 x 450 x 252.5mm [33.46 x 17.71 x 9.94in] (H x L x D) Simple fittings for 6 mm O.D. tubing or ¼" O.D. in Inlet PE-low density. ¼" OD in PHED-PTFE-SS as option Outlet Barbed stem for 12 mm (½"...
  • Page 16: Model Identification System

    Damage caused by misuse, neglect, accident or improper application or installation • Damage caused by any repair or attempted repair not authorized by Hach Company • Damage caused by the normal wear and tear of the electrodes.
  • Page 17 Specifications Damage caused by not sufficient maintenance, cleaning or calibration. • Any product not used in accordance with the instructions furnished by Hach Company •...
  • Page 18: Section 3 Analyzer Overview

    Section 3 Analyzer Overview 3.1 Overview The Polymetron 9240 Sodium Analyzer is a continuous on-line monitor for direct measurement of sodium in power generation processes. The measurement is based on a direct potentiometric technique using a highly sensitive sodium glass electrode. The difference of potential between the glass electrode and the reference...
  • Page 19: Schematic Process Overview

    Analyzer Overview 3.2 Schematic process overview The illustration below shows the major components of the analyzer. Figure 2 Working principal 1 - Sample inlet flow adjustment (one per channel) 7 - Sample level detector 13 - Sodium ion-selective electrode 2 - Fast loop sample outlet (one per channel) 8 - Drain 14 - Temperature electrode 3 - Magnetic stirrer...
  • Page 20: Presentation Of The Analyzer

    Analyzer Overview 3.3 Presentation of the analyzer 3.3.1 Analyzer front panel Figure 3 Analyzer front panel 1 - User interface 5 - Door lock 2 - Overflow vessel 6 - Reagent shelf 3 - Measuring cell 7 - Frame for panel mounting 4 - Flow rate adjustment for each channel...
  • Page 21: Analyzer Rear Panel

    Analyzer Overview 3.3.2 Analyzer rear panel Figure 4 Analyzer rear panel 1 - Local controller box 4 - Sample inlet valves 2 - Electrolyte reservoir 5 - Calibration canister 3 - Pump box (see also Pumps box on page 6 - Reactivation reagent canister...
  • Page 22: Pumps Box

    Analyzer Overview 3.3.3 Pumps box Three pumps have been factory installed in the pump box and are used for the following purposes: Figure 5 Pumps box interior Auto-calibration pump Reactivation pump Drain pump...
  • Page 23: Conditioning Reagent

    13 weeks approx. 3 weeks approx. 7 weeks pH measurement of 10 to 10.5 Note: As Hach Lange recommends DIPA to be used as the conditioning reagent, all references relating to the conditioning reagent throughout this manual will be to DIPA.
  • Page 24: Ph Regulation

    Analyzer Overview 3.5 pH regulation 3.5.1 Non-cationic applications In order to ensure the accuracy and the repeatability of low sodium concentrations, pH must be constant and preferably maintained at or above 11.2 to maintain the lowest proton interference. The 9240 uses the injection of vapor of diisopropylamine (DIPA) to obtain high pH level without sodium contamination.
  • Page 25 Analyzer Overview The usual T ratio values are as follows: water Tgas/Twater Ratio 180% Refer to Sample pH conditioning check on page 53 for the procedure to select the correct ratio in relation to the initial pH sample. DIPA consumption The consumption of DIPA will depend on the values defined in the above table.
  • Page 26: Measurement Process

    Analyzer Overview 3.6 Measurement process 3.6.1 Smart rinse option The measurement cell is rinsed by the analyzer after a calibration, grab sample, or sensor reactivation process and prior to resuming sample measurements. Set the smart rinsing parameter to No (see Measure steps on page 81) for a fixed rinse cycle of 10 minutes or set to Yes to invoke the smart rinsing option which better guarantees the accuracy of measurements...
  • Page 27: Measurement Steps

    Analyzer Overview 3.6.2 Measurement steps The 9240 is designed as a multi channel analyzer, though it can be used as a single channel analyzer. The measurement parameters are different depending on whether the analyzer is set up as a single or multi channel analyzer. If the analyzer is configured as a single channel unit, the following parameters are required: The on-line measurement time defines the measurement time of the sample, and how often •...
  • Page 28 Analyzer Overview 3.6.2.1 Search stability set to NO - Fixed measurement mode As the cycle and on-line measurement times defined by the user are fixed and identical for each channel, they need to be long enough to have an efficient rinsing step for all configurations in order to achieve accurate measurements.
  • Page 29 Analyzer Overview 3.6.2.2 Search stability set to YES - Automatic measurement mode In this configuration, a maximum cycle time not to be exceeded has to be configured. This is the sum of the maximum rinse time plus the fixed on-line measurement time. As with the fixed measurement mode, the cycle and on-line measurement times need to be long enough to meet the needs of the configuration in order to achieve accurate measurements.
  • Page 30: Measurement Cycle

    Analyzer Overview 3.6.3 Measurement cycle The following flow chart illustrates the logic used during the measurement cycle. Figure 9 Measurement cycle 3.6.4 On-line measurement On completion of each rinsing time, the analyzer starts the on-line measurement step. The sodium concentration is shown on the display panel and output via the RS485 link. At the end of this step, the analyzer refreshes the 4-20 mA outputs and alarm relays, writes data to the analyzer memory and the display panel.
  • Page 31: Automatic Calibration Cycle

    Analyzer Overview 3.7 Automatic calibration cycle The following section gives an overview of the automatic calibration process. For details on setting up and running this process, refer to the section entitled Calibration on page 93. As the 9240 is a multi channel analyzer, the user must define which channel will be used for calibration (see Automatic calibration setup on page 98).
  • Page 32 Analyzer Overview This process calculates the slope and offset of the ISE sodium electrode and the reference electrode. It is based on the measurement of the potential and temperature of three different samples, two of which contain known concentrations of sodium: Measurement of a first sample of unknown concentration •...
  • Page 33: Manual Calibration Cycle

    Analyzer Overview 3.8 Manual calibration cycle The following section gives an overview of the manual calibration process. For details on setting up and running this processes, refer to the section entitled Calibration on page 93. As the 9240 is a multi channel analyzer, the user must define which channel will be used for calibration (see Automatic calibration setup on page 98).
  • Page 34: Grab Sample Measurement Cycle

    Analyzer Overview 4. Phase 4: The overflow vessel is drained, and the system waits for the operator to fill the overflow vessel with the low value calibration solution. 5. Phase 5: The overflow vessel is drained into the measurement cell and the first concentration measured.
  • Page 35: Analyzer Outputs

    Analyzer Overview 3.10 Analyzer outputs The following tables shows the outputs (screen, analog, alarm relays and RS485) generated by the analyzer during the various processes (measurement, grab sample, calibration and sensor reactivation). For the grab sample and calibration processes, see Automatic calibration cycle on page Manual calibration cycle on page 32, and...
  • Page 36: Auto Calibration Process

    Analyzer Overview 3.10.4 Auto calibration process 4-20 mA RS485 Action Screen Alarms Conc. Other prog. Address 0 Others Phase 1 Actual Actual Actual Address 163=10 Phase 2 Address 163=13 Phase 3 Frozen Frozen Frozen Phase 4 Event Phase 5 Actual Actual Actual Calibration...
  • Page 37: Available Options

    The K-Kit option includes a gas pump with no moving parts, a power supply board, additional hydraulics and full installation instructions. For further information contact your local Hach Lange representative. 3.11.2 Static heat exchanger system A static heat exchanger system is available as an option and is capable of working with up to four channels.
  • Page 38: Section 4 Installation

    Avoid any location with a corrosive atmosphere or subject to liquid spills. • Chose a dry and dust-free location. • The ambient temperature of the analyzer should not exceed 45°C. If the temperature is • below 5°C, the analyzer should be installed in a heated cabinet (not provided by Hach Lange).
  • Page 39: Instrument Mounting

    Installation 4.3 Instrument mounting CAUTION Whether the instrument is to be mounted on a panel or wall, it is important to note that it must be placed in an upright position with the transmitter at the top. It is recommended to use a spirit level to ensure that the instrument is correctly positioned and not leaning to one side or forward.
  • Page 40: Wall Mounting

    Installation 4.3.2 Wall mounting Use the wall mounting kit to fix the instrument to the wall. Figure 14 Wall mount kit Use these to drill the four holes for fixing the instrument on the wall. The distance between the two pieces is 460 mm. CAUTION It is extremely important to respect this gap of 460 mm to avoid bending the cabinet out of shape while fitting.
  • Page 41: Step-By-Step Installation

    Installation 4.4 Step-by-step installation WARNING No intervention should be made on the instrument without first switching off the power. The electrical installation should be carried out by duly qualified personnel. A supply voltage of 100-240 VAC is acceptable without changing the configuration. The power supply terminals can be removed from their housing to make connection easier.
  • Page 42 Installation 2. Pass the power supply cable through the cable gland located at the back left of the bottom of the cabinet. 3. Open the I/O connection box on the rear panel. 4. Unscrew the two fixing elements...
  • Page 43 Installation 5..and allow the I/O box to rotate down. 6. Pass the mains power cable through the back end cable gland on the transmitter enclosure. 7. Open the transmitter front door.
  • Page 44 Installation 8. Remove the metallic shielding plate protecting access to the main board. 9. Take the power supply connector and note where the earth, live, and neutral must be connected. 10. Connect the power supply cables to the connector...
  • Page 45 Installation 11..and put the connector back in place. 12. Replace the metallic shielding plate.
  • Page 46: Rs485 Connection

    Installation 4.4.2 RS485 connection 1. Connect the RS485 communication cable as indicated. Connection is the same on the CPU board for both the JBUS/MODBUS and PROFIBUS options. 2. Close the transmitter door.
  • Page 47 Installation 3. Put the I/O box back in its normal position. 4. Fix the I/O box back in place with the 2 screws.
  • Page 48: Input/Output Connections

    Installation 4.4.3 Input/Output connections 1. Analog output – Iout 0 is used for the current signals of the measurement. Other analog outputs can be freely linked to different parameters like measurement, temperature – refer to the section entitled User Setup on page 79 for details.
  • Page 49: Sample Tubes Installation

    Installation 4.4.4 Sample tubes installation Sample inlet connections - 6 mm (or ¼'') in polyethylene or PTFE or FEP. If particulate • matter is present in the sample, pre-filtration is necessary. A filter should be inserted in the sample line. One is available as an option. Use new pipes for all connections during installation.
  • Page 50: Reagents Installation

    Installation 4.4.5 Reagents installation 1. Prepare the reagents according to Reagent Preparation on page 137. Install the reagent canisters in their place and connect them to their respective tubes. 2. Install and connect the conditioning solution (diisopropylamine). The diagrams show an analyzer for non-cationic applications.
  • Page 51: Magnetic Stirrer Installation

    Installation 4.4.6 Magnetic stirrer installation 1. On the front of the panel, remove the plastic bag from the overflow vessel 2. Remove the magnetic stirrer from the bag and install it in the overflow vessel...
  • Page 52: Reagents Volume Declaration

    Installation 4.4.7 Reagents volume declaration Note: As you will now be using the analyzer menus to input data, it may be useful to familiarize yourself with the data entry procedures by reading the section entitled Data entry on page 59 of this manual.
  • Page 53: Flow Rate Adjustment

    Installation 4.4.8 Flow rate adjustment 1. From the MAINTENANCE/DIAG. menu select START UP and press Enter 2. First, the system automatically primes both the calibration and reactivation tubes. 3. Check that there are no air bubbles in the reagent tubes for reactivation and auto calibration. 4.
  • Page 54: Sample Ph Conditioning Check

    Installation 4.4.9 Sample pH conditioning check Note: At this stage, the electrodes should not have been installed. 4.4.9.1 Non-cationic applications 1. Install a calibrated pH sensor in the center position of the measurement chamber which is normally used for the ISE sodium electrode. 2.
  • Page 55: Reference Electrode Installation

    Installation 7. For each channel, measure the pH in the conditioned sample to check if the pump ratios are efficient enough to obtain a pH of around 11.0. 8. If necessary, update the ratios of each channel to maintain a final constant pH of 11.0 ± 0.2. 4.4.10 Reference electrode installation 1.
  • Page 56 Installation 5. Install the reference electrode in the extreme left measurement chamber. 6. Connect the reference cable (the one without the blue label on it) to the reference electrode. 7. Connect the electrolyte tube to the reference electrode.
  • Page 57: Sodium Ion Selective Electrode Installation

    Installation 4.4.11 Sodium ion selective electrode installation Note: It is critical to preserve the integrity of the sodium ion selective electrode as much as possible. This is why this electrode must be installed at the very last moment after all other adjustments. 1.
  • Page 58: Fill Electrolyte Reservoir

    Installation 4.4.12 Fill electrolyte reservoir 1. The electrolyte reservoir is located at the back of the analyzer. 2. Take the KCl electrolyte bottle and insert the tip of the tapered spout into the reservoir inlet tube (illustrated right) as far as it will go but without exerting any extra pressure.
  • Page 59: Analyzer Stabilization

    Installation 4.5 Analyzer stabilization At this stage the analyzer has been completely installed, but needs to run for a period of time to stabilize. 1. On the analyzer, press Start on the main menu to start the measurement process. 2. Leave the system to run for a couple of hours before starting any calibrations. 4.6 Analyzer setup After the analyzer has stabilized, the system and user parameters must be set, followed by an initial calibration.
  • Page 60: Section 5 Operating Instructions

    Section 5 Operating Instructions Note: All screen display examples in the following sections are shown as black on a white background for reasons of clarity, and do not necessarily reflect the actual colors used on the instrument display. 5.1 Data entry 5.1.1 Function keys The display panel of the 9240 has 5 function keys (illustrated below) to allow menu option selection, field selection, and data entry options.
  • Page 61: Numeric Fields

    Operating Instructions 5.1.2 Numeric fields These fields require that the user enter one or more numeric values into a field. The type of field determines the available input. In some fields only digits 0 through 9 would be available to select whereas in other fields the decimal point and/or minus sign may also be available.
  • Page 62: Alphanumeric Fields

    Operating Instructions 5.1.3 Alphanumeric fields These fields require that the user enter one or more alphanumeric values into a field. The type of field determines the available input. In some fields only upper case alpha characters may be allowed, in others upper and lower case alphanumeric characters my be allowed, etc. In this example, a name is required to give to the channel.
  • Page 63: List Element Fields

    Operating Instructions 5.1.4 List element fields This type of data entry is where a pre-defined list of available data values are available to the user who must select the one which is applicable. Free-format text is not allowed. The highlighted field in the example left is a data element where there is only a limited number of valid options (i.e.
  • Page 64: Measurement Screens

    Operating Instructions 5.2 Measurement screens 5.2.1 Principal display Figure 17 Principal measurement screen An example of this display is given above. It shows the details of the current sample being measured. In this example, the measurement is for Channel 1 (channel name of Sample 1), but this will change as the analyzer switches to the next measurement channel in the user-defined sequence.
  • Page 65: Historical Display

    Operating Instructions 5.2.2 Historical display Figure 18 Historical measurement screen This screen shows the last sample measurements for each channel, the last grab sample measurement and the last verification and gap measurements. The example illustrated is for a 4-channel analyzer. Underneath these measurements, the actual potential between the glass electrode and the reference electrode is displayed along with the temperature.
  • Page 66: Alarms Screen

    Operating Instructions 5.2.3 Alarms screen Figure 19 Alarm screen Alarms S1-S4 relate to the four alarm outputs. This is followed by the warning alarm (W!) and the system alarm (small graphic). The message against each alarm will indicate OK (no problems encountered), INACTIVE (the alarm has been deactivated), or a message indicating the reason for the alarm.
  • Page 67: Main Menu

    Operating Instructions 5.3 Main menu For a diagrammatic view of the complete menu structure, please refer to Menu structure overview on page The main menu is accessible from any one of the measurement screens (as illustrated in Figure 17 on page 63 for example).
  • Page 68: Verification

    Operating Instructions 5.3.1 Verification This option allows you to verify the measurement using a solution of known sodium concentration. Concentration of the known solution should be higher than 20 ppb (100 ppb is recommended) to ensure a preparation step in a range with lower risks of contamination. Note: This is a measurement verification cycle only, and no adjustments to parameters or calibration data will be made.
  • Page 69 Operating Instructions Upon completion of the rinsing step, the analyzer is ready for the first manual step in the cycle. Place the sample level detector on the side and pour approximately 200 mL of the solution into the overflow vessel, as prompted on screen.
  • Page 70: Grab Sample

    Operating Instructions 5.3.2 Grab sample Note: It is recommended that the grab sample should have a concentration of > 10 ppb and should be at the same temperature as during calibration for better accuracy. Under these conditions, measurement with manual introduction gives 5% accuracy from 10 ppb to 10 ppm at a temperature which is within ± 5 °C of the temperature during calibration.
  • Page 71 Operating Instructions The display switches back to the main measurement screen and displays the measurement value. The progress bar at the top of the screen monitors the measurement time of the sample. If you wish to abort the process at any time during the measurement cycle, select Stop.
  • Page 72: Menu Structure Overview

    Operating Instructions 5.4 Menu structure overview...
  • Page 73 Operating Instructions...
  • Page 74: Section 6 System Setup

    Section 6 System Setup Before attempting to setup the analyzer, ensure that you have read and understood how to enter and update data fields as described in the section entitled Function keys on page Press the Up Arrow function key under the Menu option on the measurement screen to display the Main Menu.
  • Page 75: System Setup - Menu Overview

    System Setup 6.1 System setup - Menu overview Figure 21 System setup menu 6.2 Date and time Select the DATE/TIME option to access the date/time sub-menu. This option allows you to set up the system date and time. Scroll to the required day and press the Enter function key. Enter the day, month, and year.
  • Page 76: Display Options

    System Setup 6.3 Display options Select the DISPLAY option to set up the display parameters. Select the LANGUAGE, CONC. UNIT and TEMPERATURE UNIT fields in turn, and set your preferences by scrolling through the available options (listed in Table 3 below).
  • Page 77: Passwords

    System Setup 6.4 Passwords Select the PASSWORDS option to set passwords for access to programming, calibration and system setup options. By default all passwords are set to 0000. Each password is a 4-digit numeric field. Enter the required value for each of the three passwords. A value of 0000 signifies no password is required to gain access to those menu options.
  • Page 78: Adjust Ma Output

    ADJUST mA OUTPUT screen, to allow you to select the next output to adjust. 6.7 Factory settings This option is reserved for qualified Hach Lange service personnel. If you believe that changes to the settings may be required,...
  • Page 79 System Setup...
  • Page 80: Section 7 User Setup

    Section 7 User Setup Before attempting to setup the analyzer, ensure that you have read and understood how to enter and update data fields as described in the section entitled Function keys on page The default and available settings for these options are listed in Default Configuration on page 121.
  • Page 81: User Setup - Menu Overview

    User Setup 7.1 User setup - Menu overview Figure 22 User setup menu...
  • Page 82: Measurement

    User Setup 7.2 Measurement 7.2.1 Targeted pH (non-cationic applications only) If the analyzer has been set up for non-cationic applications, the screen illustrated left is displayed. Select the pH option and enter the target pH value for your application. The measured pH value of the sample in the measuring cell should be within ±0.2 pH of the target.
  • Page 83: Reactivation Frequency

    User Setup If the analyzer has been set up as a multi channel analyzer the screen illustrated left will be displayed. The on line measurement time is the time when the analyzer displays the true sodium measurement. This value must be at least 1 minute and less than the cycle time.
  • Page 84: Datalogger Setup

    User Setup 7.2.5 Datalogger setup This option allows you to view data held in the analyzer’s internal memory. Select the VIEW DATA option to display the requested data. All information matching the parameters selected in the VIEW SETUP option is displayed on the screen. If the data covers more than one screen, an Arrow key will be displayed at the bottom.
  • Page 85: Graph Time Base

    User Setup 7.2.6 Graph time base For graphical displays, enter the number of hours as the base line for the graph. This can be 4, 8, 12, 16, 20 or a 24 hour period. 7.3 Alarms Select the ALARMS option to set up the parameters for all the alarms including the system and warning alarms.
  • Page 86: Alarms 1 To 4

    User Setup 7.3.1 Alarms 1 to 4 Table 4 Alarms 1 to 4 parameters Value Description Trigger the alarm when the measurement is above or below a Limit pre-defined limit Sample flow Trigger the alarm when the sample flow rate is too low Mode Trigger the alarm when the channel is active i.e.
  • Page 87: Warning Alarm

    User Setup 7.3.2 Warning alarm Table 5 Warning alarm parameters Value Description Activate the warning alarm Alarm Deactivate the warning alarm Manual When the alarm is triggered, turn it off by pressing the Enter function key Accept When the alarm is triggered, it will turn itself off only when the reason for the Auto alarm being triggered is no longer valid N.O.
  • Page 88: System And Warning Alarm Table

    User Setup 7.3.4 System and warning alarm table The following table lists the different system and warning alarms: Table 7 System and warning alarms Message Description Category Reset measure mod. Reset the measure module System Reset LC module Reset the local controller module System LC Error Local controller module error...
  • Page 89: Ma Outputs

    User Setup 7.4 mA outputs Select the mA OUTPUTS option to set up the parameters for all the analog outputs. From the list available, select the mA output you wish to set. 7.4.1 Outputs 0 to 5 Table 8 Analog output parameters Description Choose the attribute that triggers the analog output: Measurement channel...
  • Page 90: Event Indication

    User Setup Figure 23 Linear and dual slopes The screen illustrated left shows an analog output to be activated on a measurement on channel 1. The lower end of the scale will be 0 mA on a value of 1 ppm and the high end of the scale will be 20 mA on a value of 12 ppm.
  • Page 91: Test

    User Setup Define the attribute for the event. This is one of the 6 mA outputs (0 to 5) or None. The MODE can be either a Preset val or Frozen. If a preset value is chosen, you will be required to enter the value of the analog output that will be forced, when that event occurs.
  • Page 92: Sample Channels

    User Setup Enter the parameters as described in Table 9 on the previous page. Press the Enter function key to accept each data element. On completion, press the Esc function key to return to the user setup screen. 7.6 Sample channels Select the SAMPLE CHANNELS option to set up the parameters for the sample channels.
  • Page 93: Channel Activation

    User Setup 7.6.2 Channel activation Using the Up and Down Arrow keys, define whether the channel is active (Activ) or inactive (Inactiv). The example shown left indicates an analyzer set up with three channels, of which channel 2 is inactive. 7.6.3 Sequence The sequence defines the channel order in which samples are measured.
  • Page 94: Section 8 Calibration

    Section 8 Calibration 8.1 General Before attempting to calibrate the analyzer, ensure that you have read and understood how to enter and update data fields as described in the section entitled Function keys on page Note: The instrument cannot be calibrated until at least one complete measurement cycle has been successfully performed.
  • Page 95: Calibration - Menu Overview

    Calibration 8.3 Calibration - Menu overview Figure 24 Calibration menu 8.4 Start calibration Select the START CALIBRATION option to calibrate the analyzer manually. 8.4.1 Calibrate known addition Select the CAL.KNOWN ADDITION option to calibrate the analyzer by mixing the sample with a known concentration of calibration solution.
  • Page 96 Calibration If the volume of solution is OK, the background point measurement (P0) to be used as the base measurement value is taken. This is the measurement against the sample before any additions of the calibration solution have been made. A progress bar is displayed showing the progress to date and approximate time remaining.
  • Page 97: One Point Calibration

    Calibration When the measurement is stable, the details are displayed at the top of the screen against measurement P1. The overflow vessel is then rinsed and re-filled with sample plus 20mL of the calibration solution. This sample is then measured and the details displayed against P2.
  • Page 98: Two Point Calibration

    Calibration When prompted, fill the overflow vessel with the calibration solution and select OK to start the measurement. The calibration solution is measured and the results displayed. Once the final measurement is stable, the offset for the calibration is calculated. The success or failure screen is then displayed (see examples in Calibration results on page...
  • Page 99: Automatic Calibration Setup

    Calibration 8.5 Automatic calibration setup Select the AUTO. CAL. SETUP option to access its sub-menu and set the calibration parameters. This allows the analyzer to be automatically calibrated at pre-defined and regular intervals. The process is the same as described in Calibrate known addition on page 94 except it is initiated automatically at the...
  • Page 100: Calibration Results

    Calibration 8.6 Calibration results After the slope and offset have been calculated, the default parameters, last calibration details and current calibration details are displayed, along with a “Calibration ok” or “Calibration error” message. The criteria for accepting or rejecting the calibration is as follows: The slope must be within ±...
  • Page 101: Custom Adjustment

    Calibration 8.8 Custom adjustment Where the analyzer displays values that are slightly above or below the expected value, the CUSTOM.ADJ option can be used to manually adjust the measurement value. Enter a positive or negative value which will be added to the measurement value.
  • Page 102: Section 9 Maintenance And Diagnostics

    (cationic applications) how often the conditioning solution will need to be changed. Solenoid valves The following items are performed by the Hach Lange service group System check-up External audit 9.2 Battery replacement The transmitter is delivered with a CR1220 (3V) lithium battery included to safeguard the time function in the instrument when not powered up.
  • Page 103: Local Controller Board Replacement

    Maintenance and Diagnostics 9.3 Local controller board replacement WARNING Potential electrocution hazard. Always disconnect power to the instrument before making any electrical connections. CAUTION Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document. If replacing the local controller board (part number 09200=A=5036) refer to Figure 25 interface connection details.
  • Page 104: Fuse Replacement

    • JDYX2, SCHURTER, Type FST, rated 2 AT, 250 VAC • If there is any difficulty in obtaining these replacement fuses, please contact your local Hach Lange representative for advice and assistance. 9.5 Cleaning and decontamination The analyzer does not normally require any cleaning or decontamination. The internal cycle is only in contact with clean, pure water.
  • Page 105: Maintenance And Diagnostics Menu Option

    Maintenance and Diagnostics 9.6 Maintenance and diagnostics menu option There are a number of maintenance and diagnostic procedures that can be performed using the software. These options can be found in the MAINTENANCE/DIAG. menu option of the main menu, and are explained in detail in the rest of this section. The maintenance and diagnostics screen gives a list of maintenance procedures.
  • Page 106: Reagent Changes

    Maintenance and Diagnostics 9.8 Reagent changes This procedure is applicable to each change of reagent solution. It allows the regulation of the volumes of solutions in the bottles. The analyzer consistently checks the consumption of these solutions, triggering an alarm when these levels get too low. This option must be selected each time a solution is refilled or changed to reset the measurement parameters.
  • Page 107: Priming Tubes

    Maintenance and Diagnostics 9.8.2 Priming tubes After refilling the bottles (reference electrolyte, calibration solution and reactivation solution) and/or after exchanging the empty bottle of conditioning reagent with a new one, select the PRIMING TUBES option. This will set off the process to purge and refill the tubes. A progress bar is displayed while the process executes.
  • Page 108: Calibration Diagnostics

    Maintenance and Diagnostics 9.9 Calibration diagnostics This option is used by after sales service technicians to verify that the last calibration was working correctly. 9.10 Raw values This option is used by after sales service technicians to verify that the electrodes are functioning correctly. 9.11 Test accessories This option allows you to check that a number of accessories, as well as the alarm relays and logical inputs are functioning...
  • Page 109: Hydraulics

    Maintenance and Diagnostics 9.11.1 Hydraulics TEST HYDRAULICS allows you to check that all accessories apart from relays and logical inputs are working correctly. The screen allows selection of the solenoid valves, the conditioning valve, the pumps, the mixer or the drain. Select the element you wish to test and then select On to activate it or Off to deactivate.
  • Page 110: Logical Inputs

    Maintenance and Diagnostics 9.11.3 Logical inputs Choose this option to check the logical inputs status. Each logical input shows either Open or Closed to indicate its current status. Press Esc to exit the screen. 9.12 Sensor reactivation Usually, the electrode is reactivated automatically based on the time delay set in Reactivation frequency on page This option allows for a manual reactivation.
  • Page 111: Extended Stop

    Maintenance and Diagnostics 9.13 Extended stop If the instrument is not to be used for an extended period of time, select this option to shut the analyzer down in a controlled manner. Place all the tubes in demineralized water for cleaning. Press OK to continue A progress bar will be displayed whilst the shut down is being performed.
  • Page 112: Startup

    Maintenance and Diagnostics 9.14 Startup This process guides you through various steps required to set the instrument up for initial measurements, or to restart after a long period of inactivity. This includes regulating the sample flow and purging the pumps of reagent. This step allows you to regulate the sample flow on each configured measurement channel.
  • Page 113 Maintenance and Diagnostics...
  • Page 114: Section 10 Troubleshooting

    Section 10 Troubleshooting 10.1 General faults WARNING Before opening the instrument, ensure the power supply has been switched off. Table 11 General Faults MALFUNCTION POSSIBLE CAUSE REMEDIES The instrument is not properly connected. No power. Check power and connection. Replace fuse (see Fuse replacement on Defective fuse.
  • Page 115: Troubleshooting

    Troubleshooting Table 11 General Faults MALFUNCTION POSSIBLE CAUSE REMEDIES Calibration is not correct. See below. Activation has been skipped. Start a new grab sample again. Set activation setup to 24 hours frequency. Grab sample is not correct. Manually activate the electrode by immersing it Electrode too slow.
  • Page 116 Troubleshooting Table 11 General Faults MALFUNCTION POSSIBLE CAUSE REMEDIES Option not activated. Call a Service Technician. Verify the electrical connections and voltage. If Pump is not connected. the voltage is around 70 volts change the Pump not working pump. (cationic applications). Verify the instrument is in START mode.
  • Page 117: Detection Of Functional Faults

    Troubleshooting 10.2 Detection of functional faults Table 12 Lack of precision PROBLEM CAUSE/SOLUTION Electrolyte is polluted. Check parts and change the electrolyte. Electrolyte leaks. Check parts and change the electrolyte. Calibrate again to check the parameters. If the error is Error during calibration or incorrect calibration.
  • Page 118: Section 11 Spare Parts And Accessories

    Section 11 Spare Parts and Accessories 11.1 Accessories - Options - Maintenance kits Table 14 Accessories, options & maintenance kits Description Part Number Optional kit for Profibus DP: board + operator manual 09125=A=1485 Optional RS485 JBUS/MODBUS: board + instruction manual (5 languages) 09125=A=2485 1 year spare part kit for 9245-9240 (all ranges).
  • Page 119: Spare Parts - In Contact With Sample

    Spare Parts and Accessories 11.2 Spare parts - In contact with sample Table 15 Spare parts - In contact with sample Description Part Number Cleaning brush for Static heat exchanger system 09240=A=9840 Sample level detector kit (cable, level detector, overflow cell cap, update instructions) 09240=A=0260 ø...
  • Page 120: Spare Parts - In Contact With Reagents Or Standard Solution

    Spare Parts and Accessories 11.4 Spare parts - In contact with reagents or standard solution Table 17 Spare parts - In contact with reagents or standard solution Description Part Number Porous cartridge for gaseous conditioning in 9245-9240 09073=C=0340 DIPA Merck bottle cap adapter (GL40 -> GL45) 09073=C=0350 Orion monoethylamine bottle cap adapter 09186=C=0360...
  • Page 121: Additional Hardware

    Spare Parts and Accessories 11.6 Additional hardware Table 19 Additional hardware Description Part Number Locking key for 924x enclosure 32965 Elbow torque screw driver T6 820=000=006...
  • Page 122: Section 12 Default Configuration

    Section 12 Default Configuration 12.1 User configuration table User Configuration Option Default Value Possible Values Selected Values Date MEASUREMENT 11.20 10.00 - 12.00 Tgaz/Twater % 0000 - 0250 On-line measure time 10 minutes 0 - 9999 minutes Smart rinsing Yes or No Max.
  • Page 123: Default Configuration

    Default Configuration User Configuration Option Default Value Possible Values Selected Values Date ALARMS / ALARM 3 Mode Limit Limit Active channel Sample flow None Attrib Channel 1 Channel 1 - 4 Limit 1.00 ppm 0 - 9990 ppm Direction Down Delay 0 seconds 0 - 999 seconds...
  • Page 124 Default Configuration User Configuration Option Default Value Possible Values Selected Values Date mA OUTPUTS / OUTPUT 1 Attrib Channel 1 Channel 1 - 4 Temper. None Type 0-20 mA 0-20 mA 4-20 mA Mode Linear Linear Dual Logarithm 0 ppb 0 - 9990 ppm Middle 100 ppb...
  • Page 125 Default Configuration User Configuration Option Default Value Possible Values Selected Values Date mA OUTPUTS / OUTPUT 4 Attrib None Channel 1 - 4 Temper. None Type 0-20 mA 0-20 mA 4-20 mA Mode Linear Linear Dual Logarithm 0 ppb 0 - 9990 ppm Middle 100 ppb 0 - 9990 ppm...
  • Page 126 Default Configuration User Configuration Option Default Value Possible Values Selected Values Date mA OUTPUTS / EVENT INDICATION / WARNING ALARM Attrib None None Output 0 - 5 Mode Frozen Preset value Frozen mA OUTPUTS / EVENT INDICATION / SYSTEM ALARM Attrib None None...
  • Page 127 Default Configuration...
  • Page 128: Section 13 Material Safety Data Sheets (Msds)

    Section 13 Material Safety Data Sheets (MSDS) 13.1 Diisopropylamine (DIPA) 1. IDENTIFICATION OF THE SUBSTANCE Catalogue No: 803646 Product name: diisopropylamine for synthesis 2. COMPOSITION / INFORMATION ON INGREDIENTS CAS no: 108-18-9 Molecular weight: 101.19 Chemical formula: C EC index no: 612-048-00-5 EINECS number: 203-558-5 3.
  • Page 129 Material Safety Data Sheets (MSDS) 9. PHYSICAL AND CHEMICAL PROPERTIES Form: liquid Color: colorless Odor: amine-like Ph value: not available Melting temperature: -96°C Boiling temperature: 83 - 84°C Ignition temperature: 295°C DIN51794 Flash point: -17°C DIN51755 Explosion limits: Lower: 1.5 vol% •...
  • Page 130 Material Safety Data Sheets (MSDS) 14. TRANSPORT INFORMATION DOT: Shipping Name: Diisopropylamine • UN Number: UN1158 • Hazard Class: 3,8 • Packing Group: PG II • Label(s): Flammable liquid, corrosive • IATA: Shipping Name: Diisopropylamine • UN Number: UN1158 • Hazard Class: 3,8 •...
  • Page 131: Potassium Chloride

    Material Safety Data Sheets (MSDS) 13.2 Potassium chloride 1. IDENTIFICATION OF THE SUBSTANCE Intended/recommended use: electrolyte for reference electrode in 9245 sodium analyzer Chemical formula: KCl 3M in water Molecular weight: 74.55 g/mol Product name: Potassium Chloride 2. COMPOSITION / INFORMATION ON INGREDIENTS Ingredients: Hazardous KCl:...
  • Page 132 Material Safety Data Sheets (MSDS) 8. EXPOSURE CONTROLS AND PERSONAL PROTECTION Protective clothing: None: Overalls: Glasses: Gloves: X Other: Ventilation: None: Vent hood: Mechanical (General): Special: Individual protection: Protective clothing should be selected specifically for the working place. Wash hands after handling. Avoid prolonged exposure.
  • Page 133: Sodium Chloride

    Material Safety Data Sheets (MSDS) 13.3 Sodium chloride 1. IDENTIFICATION OF THE SUBSTANCE Intended/recommended use: 10 ppm (Na) calibration standard for 9245 sodium analyzer Chemical formula: NaCl 10 ppm in water Molecular weight: 58.44 g/mol Product name: Sodium Chloride 2. COMPOSITION / INFORMATION ON INGREDIENTS Ingredients: Hazardous NaCl:...
  • Page 134 Material Safety Data Sheets (MSDS) 8. EXPOSURE CONTROLS AND PERSONAL PROTECTION Protective clothing: None: X Overalls: Glasses: Gloves: Other: Ventilation: None: X Vent hood: Mechanical (General): Special: 9. PHYSICAL AND CHEMICAL PROPERTIES Appearance: liquid. Odor: odorless. Color: colorless Boiling point: 100°C Vapor pressure: Vapor density (air=1): N/A Melting point: 0°C...
  • Page 135: Sodium Nitrate

    Material Safety Data Sheets (MSDS) 13.4 Sodium nitrate 1. IDENTIFICATION OF THE SUBSTANCE Intended/recommended use: 0.5 M NaNO reactivation solution for 9245 sodium analyzer Chemical formula: NaNO 0.5 M (42.5 g/L) in water Molecular weight: 85 g/mol Product name: Sodium Nitrate 2.
  • Page 136 Material Safety Data Sheets (MSDS) 8. EXPOSURE CONTROLS AND PERSONAL PROTECTION Protective clothing: None: X Overalls: Glasses: Gloves: X Other: Ventilation: None: X Vent hood: Mechanical (General): Special: 9. PHYSICAL AND CHEMICAL PROPERTIES Appearance: liquid. Odor: odorless. Color: colorless Boiling point: 100-103°C Vapor pressure: Vapor density (air=1): N/A Melting point: 0°C...
  • Page 137 Material Safety Data Sheets (MSDS)
  • Page 138: Section 14 Reagent Preparation

    Protective eye wear is always recommended when contact with chemicals is possible. 14.1 Conditioning Reagent Use of Diisopropylamine in the conditioning bottle is the Hach Lange recommended solution • (CAS no: 108-18-9; see Diisopropylamine (DIPA) on page 127 for the MSDS).
  • Page 139: Automatic Calibration Solution (10 Ppm Na)

    Reagent Preparation Standard Solution 100 ppb This solution is of LOW concentration for manual calibration and verification samples. 1. Rinse at least 3 times a 1L volumetric flask (“A” class) with ultra pure water. The cleanliness of the flask is essential. 2.
  • Page 140: Kcl

    Reagent Preparation 14.4 3M KCl 1. To prepare 1L of 3M KCl, rinse at least 3 times a 1L volumetric flask (“A” class) with ultra pure water. The cleanliness of the flask is essential. 2. Put 223.5 g of KCl inside this flask and add ultra pure water almost to the line delimiting the volume.
  • Page 141 Reagent Preparation...
  • Page 142 Tel. +49 (0) 2 11 52 88-320 SWITZERLAND Fax (970) 669-2932 Fax +49 (0) 2 11 52 88-210 Tel. +41 22 594 6400 orders@hach.com info-de@hach.com Fax +41 22 594 6499 www.hach.com www.de.hach.com © Hach Company/Hach Lange GmbH, 2010–2016. All rights reserved.

Table of Contents