Standard Solution Method
Prepare a 0.60 mg/L NH
–N standard by pipetting 6.0 mL of a 10 mg/L Ammonia
3
Nitrogen Standard into a 100-mL volumetric flask. Dilute to 100 mL with
ammonia-free water. Use 10 mL of this 0.60 mg/L ammonia nitrogen standard
directly in step 3 of the procedure. Perform the procedure as described.

Method Performance

Precision
Typical Precision (95% Confidence Interval) = 0.60 ± 0.05 mg/L NH
–N
3
Estimated Detection Limit
EDL = 0.01 mg/L NH
–N
3

Standard Calibration Adjust Method

To perform a standard calibration adjustment using the 0.60 mg/L ammonia–
nitrogen standard or using an alternative concentration, see
Standard Calibration
Adjust on page
2—13.