Chapter 4
Calibrating
Calibrating the mass scale entails setting up a calibration file and specifying calibration
parameters in MassLynx.

4.1 Setting Up the Calibration File

Before calibrating, you must remove the current calibration file and select a reference file:
1. Select
Calibration > Calibrate Instrument
Calibration window appears
Reference File
List
2. Select
NaICs2
Mass Detector is a 2000 model. Select
(Figure
4-1).
Figure 4-1 Calibration Window
from the reference file list to choose the reference file, if your ZQ
from the Tune window
NaICs4
if it is a 4000 model.
Setting Up the Calibration File
(Figure
3-3). The
54