2.1 Site Selection and Power Requirements

Install the detector on a stable, level, and appropriately clean surface that meets the
specifications in
Table 2-1 Installation Site Requirements
Factor
Temperature range
Relative humidity
range
Bench space
Clearance
Power requirements
Electromagnetic fields No nearby source of electromagnetic noise, such as NMR
Static electricity
Vibration
Attention:
To avoid overheating the instrument, allow the prescribed clearances at the
STOP
top, sides, and rear of the instrument.
Attention:
To avoid damaging the instrument, measure the voltage at the 230 VAC outlet
STOP
before installing the instrument. Readings of less than 208 VAC indicate the need for a
step-up transformer.
Table
2-1.
15 to 28 °C (59 to 82.4 °F)
20 to 80%, noncondensing
Width: 15.2 in. (38.1 cm)
Depth: 26 in. (66.1 cm)
Height: 23 in. (57.2 cm)
Weight: 210 lb. (95.3 kg)
Rear: 4.75 in. (120 mm)
Right side: 20 in. (0.5 m) to allow for service access.
Note:
Movable equipment can be located as close as 4.75 in.
(120 mm).
Left side: 0.0 in./mm
Top: 11 in. (28 cm)
Grounded AC, 230 V, 50/60 Hz
systems or magnetic sector mass spectrometers
Negligible
Negligible
Requirement
2
Installing
31