Parameter
BYPASS
0
SHED_OPT
0
HI_HI_LIM
+INF
HI_LIM
+INF
LO_LIM
-INF
LO_LO_LIM
-INF
OUT_HI_LIM
100
OUT_LO_LIM
0
SP_HI_LIM
100
SP_LO_LIM
0
Revision 2.0
FOUNDATION Fieldbus ST 700 Pressure Transmitter User's Guide
Initial Value
Valid Range
1:OFF, 2:ON
1-8 see Shed
Options in the FF
specs.)
PV_SCALE, +INF
PV_SCALE, -INF
OUT_SCALE
+/- 10%
PV_SCALE
+/- 10%
Corrective Action
Initial value is a configuration error.
Set value in valid range.
Initial value is a configuration error.
Set value in valid range.
Values must be set in rank order. For
example, LO_LIM > LO_LO_LIM but <
HI_LIM etc.
Values must be set in rank order.
Verify that OUT_HI_LIM > OUT_LO_LIM.
Verify that SP_HI_LIM > SP_LO_LIM.