Setting the Temperature

Section 8.2 explains in detail how to set the temperature set-point on the cali-
brator using the front panel keys. The procedure is summarized here.
1. Press "SET" twice to access the set-point value.
2. Press "UP" or "DOWN" to change the set-point value.
3. Press "SET" to program in the new set-point.
4. Press "EXIT" to return to the temperature display.
When the set-point temperature is changed the controller switches the well
heater on or off to raise or lower the temperature. The displayed well tempera-
ture gradually changes until it reaches the set-point temperature. The well may
require 5 to 10 minutes to reach the set-point depending on the span. Another 5
to 10 minutes is required to stabilize within ±0.1°C of the set-point. Ultimate
stability may take 15 to 20 minutes more of stabilization time.