9132 Portable IR Calibrator
User's Guide
7.13.1
7.13.2
7.13.3
30
rameter to the new value. Pressing "EXIT" causes the parameter to be skipped
ignoring any changes that may have been made.
R0
This probe parameter refers to the resistance of the control probe at 0°C. The
value of this parameter is set at the factory for best instrument accuracy.

ALPHA

This probe parameter refers to the average sensitivity of the probe between 0
and 100°C. The value of this parameter is set at the factory for best instrument
accuracy.

DELTA

This probe parameter characterizes the curvature of the resistance-temperature
relationship of the sensor. The value of this parameter is set at the factory for
best instrument accuracy.