Interface Commands - Fluke 9132 User Manual

Portable ir calibrator
Hide thumbs Also See for 9132:
Table of Contents

Advertisement

8.3
Table 2 Controller Communications Commands
Command Description
Display Temperature
Read current set-point
Set current set-point to n
Read temperature
Read temperature units
Set temperature units:
Set temperature units to Celsius
Set temperature units to
Fahrenheit
Read scan mode
Set scan mode
Read scan rate
Set scan rate
Secondary Menu
Read proportional band setting
temperature set-point and view or program the various parameters. The inter-
face commands are discussed in Section 8.3. All commands are ASCII charac-
ter strings terminated with a carriage-return character (CR, ASCII 13).

Interface Commands

The various commands for accessing the calibrator functions via the digital in-
terfaces are listed in this section (see Table 2). These commands are used with
the RS-232 serial interface. The commands are terminated with a carriage-re-
turn character. The interface makes no distinction between upper and lower
case letters, hence either may be used. Commands may be abbreviated to the
minimum number of letters, which determines a unique command. A command
may be used to either set a parameter or display a parameter depending on
whether or not a value is sent with the command following a "=". For example,
"s=150.0" sets the set-point to 150.0 degrees.
In the following list of commands, characters or data within brackets, "[" and
"]" are optional for the command. A slash, "/", denotes alternate characters or
data. Numeric data, denoted by "n", may be entered in decimal or exponential
notation. Characters are shown in lower case although upper case may be used.
Spaces may be added within command strings and will simply be ignored.
Backspace (BS, SCII 8) may be used to erase the previous character. A termi-
nating CR is implied with all commands.
Command
Format
s[etpoint]
s[etpoint]=n
t[emperature]
u[nits]
u[nits]=c/f
u[nits]=c
u[nits]=f
sc[an]
sc[an]=on/off
sr[ate]
sr[ate]=n
pr[opband]
Command
Example
Returned
s
set: 999.9 {C or F}
s=200.0
t
t: 999.9 {C or F}
u
u: x
u=c
u=f
sc
scan: {ON or OFF}
sc=on
sr
srat: 99.9 {C or F}/min
sr=1.1
pr
pb: 999.9
8 Digital Communication Interface
Interface Commands
Returned
Acceptable
Example
Values
set: 100.00 C
Instrument
Range
t: 55.6 C
u: C
C or F
scan:ON
ON or OFF
srat:12.4C/min
.1 to 99.9
pb: 15.9
33