7103 Micro-Bath User's Guide

Power

Turn on the power to the calibrator by toggling the switch on the power entry module.
The fan should begin blowing air through the instrument and the con troller display
should illuminate after 3 seconds. After a brief self test the con troller should begin
normal operation. If the unit fails to operate please check the power connection.
The display shows the well temperature and the well TEDs start operating to bring
the temperature of the well to the set-point temperature. Insure that the fluid is being
stirred.
4.3
Power
Plug the Micro-Bath power cord into a mains outlet of the proper voltage, fre quency,
and current capability. Refer to Section 3.1, Specifications, for power details.
Turn the bath on using the rear panel "POWER" switch. The Micro-Bath will turn on
and begin to heat to the previously programmed temperature set-point. The front panel
LED display will indicate the actual bath temperature.
4.4

Setting the Temperature

Section 9.2 explains in detail how to set the temperature set-point on the cali brator
using the front panel keys. The procedure is summarized here.
1.
2.
3.
4.
When the set-point temperature is changed the controller switchs the well heater on
or off to raise or lower the temperature. The displayed well tempera ture gradually
changes until it reaches the set-point temperature. The well may require 90 minutes
to reach the set-point depending on the span. Another 10 to 15 minutes is required to
stabilize within ±0.04°C of the set-point. Ultimate sta bility may take 20 to 30 minutes
more of stabilization time.
12
Shop for Fluke products online at:
Press "SET" twice to access the set-point value.
Press "UP" or "DOWN" to change the set-point value.
Press "SET" to store the new set-point.
Press "EXIT" to return to the temperature display.