Quick Start

Unpacking

Unpack the calibrator carefully and inspect it for any damage that may have occurred during
shipment. If there is shipping damage, notify the carrier immediately.
Verify that the following components are present:
9132 Calibrator
Power Cord
User's Guide
Serial Cable
9930 Software

Set Up

Place the calibrator on a flat surface with at least 6 inches of free space around the instrument.
The prop may be swung down to raise the front of the instrument from a horizontal position.
Plug the power cord into a grounded mains outlet. Observe that the nominal voltage corre-
sponds to that indicated on the back of the calibrator.
Turn on the power to the calibrator by toggling the power switch on. The fan should begin
quietly blowing air through the instrument and the controller display should illuminate after 3
seconds. After a brief self-test the controller should begin normal operation. If the unit fails to
operate please check the power connection.
The heater will start operating to bring the temperature of the calibrator to the set-point tem-
perature and the display will begin to show the actual target temperature.

Power

Plug the dry-well power cord into a mains outlet of the proper voltage, frequency, and current
capability. Refer to Specifications for power details. Turn the dry-well on using the rear panel
"POWER" switch. The dry-well will turn on and begin to heat to the previously programmed
temperature set-point. The front panel LED display will indicate the actual dry-well tempera-
ture.

Setting the Temperature

Section 8.2 explains in detail how to set the temperature set-point on the calibrator using the
front panel keys. The procedure is summarized here.
1.
Press "SET" twice to access the set-point value.
2.
Press "UP" or "DOWN" to change the set-point value.
3.
Press "SET" to program in the new set-point.
4.
Press "EXIT" to return to the temperature display.