2.10.47 Instrument's Module Configuration Information (_MDS or MDR)

IM 04L51B01-17EN
The output in response to the command "_MDS" or "_MDR" is shown below.
Syntax
EA<crlf>
p1,p2,p3,p4,p5,p6,p7,p8,p9,p10,p11<crlf>
p1,p2,p3,p4,p5,p6,p7,p8,p9,p10,p11<crlf>
...
EN<crlf>
Output Example
EA<crlf>
Main,0,1,'GX90YD-06-11',1234567,R1.01.01,,0,0,6,----------------
<crlf>
Main,0,9,'GX90EX-02-ET1'1234567,R1.01.01,,0,0,0,----------------
<crlf>
Sub,1,0,'GX90XA-10-U2',1234567,R1.01.01,,0,10,0,----------------
<crlf>
Sub,1,1,'GX90XA-10-U2',1234567,R1.01.01,,0,10,0,----------------
<crlf>
Sub,1,2,'GX90XA-10-U2',1234567,R1.02.01,,0,10,0,----------------
<crlf>
Sub,2,0,'GX90XA-10-U2',1234567,R1.02.01,,0,10,0,----------------
<crlf>
Sub,2,1,'GX90XD-16-11',1234567,R1.01.01,,0,16,0,----------------
<crlf>
EN<crlf>
One line (p1 to p11) contains configuration information of a single module.
p
Value
Description
n
Unit dependency (main or sub) information.
p1
Main, Sub
Main: Main unit (Only a single one exists in a system. GX/GP or
GM main unit)
Sub: Sub unit (Units other than the main unit. GX/GP
expandable I/O or GM sub unit)
Address number of the unit that the module is installed in. Fixed
p2
0, 1, 2
at 0.
Slot number of the unit that the module is installed in (0
p3
0, 1, 2
reference).
P4
'GX90YD-06-11',
Module name (model name). Enclosed in single quotation
marks.
'GX90EX-02-ET1',
'GX90XA-10-U2',
'GX90XD-16-11'
p5
1234567
Product serial number.
p6
R1.01.01,
Module firmware version. The output format is "R+version."
R1.02.01
p7
Space
Options. Codes of installed options delimited by spaces.
p8
0
Fixed at 0.
p9
0, 10, 8
Maximum number of input channels allowed on the module. If
there are no inputs, 0 is output.
p10 0, 16
Maximum number of output channels allowed on the module. If
there are no outputs, 0 is output.
p11 ---------------- Module status. The
string.
All modules installed in the main unit
A module installed in an expandable I/O or sub unit
(excluding the expansion module)
Module status
2.10 ASCII Output Format
is output in a character
2-175