2.9 Responses to Commands
2.9.3

Data Output Response

ASCII Output
2-118
Response Example 3
In a string of two commands, if error number 10 occurs in the second parameter of the first
command and error number 500 occurs in the fifth parameter of the second command, the
recorder outputs:
E1,10:1:2,500:2:5
Error Messages
You can use the "instrument's error number information output command" (_ERR) to output
the error message that corresponds to an error number of a negative response.
There are two types of data output: ASCII and binary.
The responses to the following commands are in ASCII.
• Queries for operation commands and setting commands
• ASCII data output requests of output commands
Syntax
EACRLF
ASCII string data • • • • • • • • CRLF
ASCII string data • • • • • • • • CRLF
|
ASCII string data • • • • • • • • CRLF
ENCRLF
The recorder adds a header (EA) in front of the ASCII string output data and a footer (EN)
at the end. The recorder adds the two characters CRLF to the end of headers, footers, and
ASCII string data.