Sensor Accuracy Guide - Welch Allyn Spot Vital Signs 420 Series Operator's Manual

Table of Contents

Advertisement

Sensor Accuracy Guide
Sensor Models
SpO; Range
70% - 100%
OXIMAX Sensor Models Single Patient Use
MAX-AI,* MAX-ALI*
+2
MAX-PI*
+2
MAX-11*
+2
MAX-RI**
+ 3.5
OxiCliq Sensor Models Single Patient Use
OXICLIO-A
+ 2.5
OXICLIQ-P
+265
Reusable Sensor Models
D-YS (Infant to Adult)
+3
D-YS & D-YSE
+ 3.5
DS-100A
+3
OXI-A/N (Adult)
+3
ΟΧΙ-Ρ/
+3
* The accuracy specification under motion conditions is + 3. For a definition of motion, contact
Nellcor Technical Services or your local Nellcor representative.
**The accuracy specification has been determined between saturations of 80% - 100%.
Accuracy Specifications: Accuracy specifications are based on controlled hypoxia studies with
healthy, non-smoking adult volunteers over the specified saturation SpO2 range. Pulse oximeter
SpO2 readings were compared to Sa02 values of drawn blood samples measured by
hemoximetry. All accuracies are expressed as + "X" digits. This variation equals + one standard
deviation (4 1 SD), which encompasses 68% of the population.