Connecting The Cables; Arm Cable; Mains Cable (Concealed Wiring) - Planmeca intra Installation Manual

X-ray unit
Hide thumbs Also See for intra:
Table of Contents

Advertisement

STANDARD INSTALLATION TO A WALL

4.6 Connecting the cables

Arm cable

Connect the grounding lead of the arm cable to the grounding point located on the right side of the
generator assembly. Connect the control leads of the arm cable (6-pole connector) to the terminal
P8 on the generator PCB. Connect the power leads (3-pole connector) of the arm cable to the termi-
nal P1 on the generator PCB. Snap the ferrite to the arm cable as shown on the figure on next page.

Mains cable (concealed wiring)

In the case that mains voltage is supplied via concealed wiring route the mains cable between the
adapter plate and the generator assembly.
Connect the grounding lead to the grounding point next to the mains switch. Connect the neutral
wire to the mains input terminal (P5) marked N. Connect the live wire to the mains input terminal
(P5) marked L.
Generator PCB version -C or later (from X-ray unit's serial number IXRF58018): When the con-
cealed wiring is used the cable coming from the ON/OFF switch to the connector P7 must be moved
to the connecto P10. Also the fuse F3 must be removed.
20 Planmeca Intra X-ray unit
WARNING
Ensure that the power supply is switched off before connecting the cables.
Neutral
Line
Blue
Brown
Blue
Brown
Remove Fuse F3
Installation manual