(2) EMEM Dialog Box Does Not Appear at Debugger Startup
Table 5.2 Checkpoints of errors at debugger startup
Error
Communication error occurred.
Data was not sent to the target.
User system cannot be properly built.
M3T-PD308F version is not the same version as
the firmware in the target.
Target MCU is in the reset state.
Target MCU cannot be reset.
Target is in "HOLD" state.
Target clock is stopped.
Target MCU is not receiving power.
REJ10J1005-0200 Rev.2.00 April 1, 2007
Check all emulator debugger settings, interface cable settings and
switches on the rear of the PC7501 match.
See the user's manuals of PC7501 and emulator debugger.
(1) Download the proper firmware.
See "2.7 Downloading Firmware" (page 26).
(2) Recheck the connection between the PC7501 and this product.
See "2.4 Connecting the PC7501" (page 22).
Download the proper firmware.
See "2.7 Downloading Firmware" (page 26).
(1) Check the reset pin of the user system is pulled up.
(2) Check the reset pin of the user system has changed from "L" to "H"
level.
(1) Check pin NMI# is held high.
(2) If in memory expansion mode or microprocessor mode, check pins
RDY# and HOLD# are held high.
(3) If the reset circuit of the user system has a watchdog timer, disable
the watchdog timer.
(4) Check that power is properly supplied to the user system and that the
user system is properly grounded.
(5) The program may be uncontrollable in areas where memory is not
allocated. Recheck the map setting.
(1) The MCU is either in stop mode or wait mode. Either reset the MCU
or cancel the mode with an interrupt.
See MCU specifications.
(2) The program may be uncontrollable in areas where memory is not
allocated. Recheck the map setting.
(1) When the clock is supplied from an external oscillator, check that the
oscillator circuit in the user system is oscillating properly.
(2) Check the switches in the emulation probe are correctly set.
See "2.10.1 Setting Switches of Emulation Probe" (page 36).
Check that power is properly supplied to the user system and that the
user system is properly grounded.
5. Troubleshooting