14-6
SCPI Reference Tables
Table 14-4 (continued)
SENSe command summary
Command
:AUTO?
:ULIMit <NRf>
:ULIMit?
:LLIMit <NRf>
:LLIMit?
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
:ADVanced
:NTOLerance <n>
:NTOLerance?
[:STATe] <b>
[:STATe]?
[:STATe] <b>
[:STATe]?
:MEDian
:RANK <NRf>
:RANK?
[:STATe] <b>
[:STATe]?
Note: *RST default is ON and SYSTem:PRESet default is OFF.
Table 14-5
STATus command summary
Command
:STATus
:MEASurement
[:EVENt]?
:ENABle <NDN> or <NRf>
Description
Query state of autorange.
Select autorange upper limit; -0.021 to
0.021 (amps).
Query upper limit for autorange.
Select autorange lower limit; -0.021 to 0.021
(amps).
Query lower limit for autorange.
Path to control the Digital Filter:
Select filter control; MOVing or REPeat.
Query filter control.
Specify filter count; 2 to 100.
Query filter count.
Advanced filter:
Specify noise tolerance (in %); 0 to 105.
Query noise tolerance.
Enable or disable advanced filter.
Query state of advanced filter.
Enable or disable digital filter.
Query state of digital filter.
Path to control median filter:
Specify "n" for rank; 1 to 5 (rank = 2n+1).
Query rank.
Enable or disable median filter.
Query state of median filter.
Description
Measurement event registers:
Read the event register.
Program the enable register.
Model 6485 Picoammeter Instruction Manual
Default
parameter
2.1e-2
2.1e-9
MOV
10
0
OFF
OFF
1
OFF
Default
parameter
(Note 1)
(Note 2)
(Note 3)
Ref.
Section SCPI
4
4
Ref
Section SCPI
10