Keithley 6485 Instruction Manual page 186

Picoammeter
Hide thumbs Also See for 6485:
Table of Contents

Advertisement

14-8
SCPI Reference Tables
Table 14-6
SYSTem command summary (see
Command
:SYSTem
:ZCHeck
[:STATe] <b>
[:STATe]?
:ZCORrect
[:STATe] <b>
[:STATe]?
:ACQuire
:PRESet
:LFRequency <freq>
:LFRequency?
:AUTO
:SYSTem:
:LFRequency
:AUTO (continued)
[:STATe] <b>
[:STATe]?
:AZERo
[:STATe] <b>
[:STATe]?
:TIME
:RESet
:POSetup <name>
:POSetup?
:VERSion?
:ERRor
[:NEXT]?
:ALL?
:COUNt?
:CODE
[:NEXT]?
:ALL?
Section 13
for detailed information)
Description
Zero check:
Enable or disable zero check.
Query state of zero check.
Zero correct:
Enable or disable zero correct.
Query state of zero correct.
Acquire a new zero correct value.
Return to SYSTem:PRESet defaults.
Select power line frequency; 50 or 60 (Hz).
Query frequency setting.
Path to control auto frequency.
Enable or disable auto frequency.
Query state of auto frequency.
Path to control autozero:
Enable or disable autozero.
Query state of autozero.
Timestamp:
Reset timestamp to 0 seconds.
Select power-on setup; RST, PRESet, or SAVx
(where x = 0 – 2).
Query power-on setup.
Query SCPI revision level.
Read error queue:
Read and clear oldest error/status (code and
message).
Read and clear all errors/status (code and
message).
Read the number of messages in queue.
Code numbers only:
Read and clear oldest error/status (code only).
Read and clear all errors/status (codes only).
Model 6485 Picoammeter Instruction Manual
Default
parameter
ON
OFF
60
ON
ON
PRES
(see Note)
Ref
Section SCPI
13
2
2
1
2
10

Advertisement

Table of Contents
loading

Table of Contents