Keithley 6485 Instruction Manual page 162

Picoammeter
Hide thumbs Also See for 6485:
Table of Contents

Advertisement

11-4
Common Commands
C) SAV <NRf> — save
RCL <NRf> — recall
Parameters
Use the *SAV command to save the present instrument setup configuration in memory for
later recall. Any control affected by *RST can be saved by the *SAV command. The *RCL
command is used to restore the instrument to the saved setup configuration. Three setup
configurations can be saved and recalled.
Model 6485 ships from the factory with SYSTem:PRESet defaults loaded into the avail-
able setup memory. If a recall error occurs, the setup memory defaults to the SYS-
Tem:PRESet values.
Programming example:
*SAV 2
*RST
*RCL 2
D) RST — reset
When the *RST command is sent, Model 6485 performs the following operations:
1.
2.
3.
E) TRG — trigger
Use the *TRG command to issue a GPIB trigger to Model 6485. It has the same effect as a
group execute trigger (GET).
Use the *TRG command as an event to control operation. Model 6485 reacts to this trigger
if BUS is the programmed arm control source. The control source is programmed from the
TRIGger subsystem.
NOTE
0 = Memory location 0
1 = Memory location 1
2 = Memory location 2
' Save present setup in memory location 2.
' Set 6485 to RST defaults.
' Return (recall) 6485 to setup stored in memory location 2.
Returns Model 6485 to the RST default conditions (see "Default" column of SCPI
tables).
Cancels all pending commands.
Cancels response to any previously received *OPC and *OPC? commands.
Details on triggering are covered in
Model 6485 Picoammeter Instruction Manual
Save present setup in memory
Return to setup stored in memory
Return Model 6485 to RST defaults
Send bus trigger to Model 6485
Section
7.