Model 6485 Picoammeter Instruction Manual
A) SYSTem:PRESet
Returns the instrument to states optimized for front panel operation. SYSTem:PRESet
defaults are listed in the SCPI tables in
B) SYSTem:TIME:RESet
Resets the absolute timestamp to 0 seconds. The timestamp also resets when power is
cycled, or after the instrument is on for 99,999.99 seconds. The TRACe:TSTamp:FORMat
command is used to select the absolute timestamp. See
C) SYSTem:POSetup <name>
Parameters
The *RST and SYSTem:DEFaults are listed in the SCPI tables in the
is saved in memory using the *SAV command. See
details.
D) SYSTem:VERSion
Read the version of the SCPI standard being used by Model 6485. Example response mes-
sage: 1996.0.
E) SYSTem:KEY <NRf>
Parameters
This command is used to simulate front panel key presses. For example, send the follow-
ing command to simulate pressing the "MEDN" key: SYSTem:KEY 2. The key-press
codes are also shown in
Section
RST
Power-up to *RST defaults
PRESet
Power-up to SYSTem:PRESet defaults
SAVx
Power-up to setup stored in memory (x = memory location
0, 1 or 2)
1 = CONFIG/LOCAL key
2 = MEDN key
3 = AVG key
4 = MX + B key
5 = M/X + B key
6 = LOG key
7 = REL key
8 = ZCHK key
9 = -------
10 = -------
11 = RANGE up arrow key
12 = AUTO key
13 = RANGE down arrow key
14 = ENTER key
15 = Cursor right arrow key
16 = ZCOR key
Table
13-3.
DISPlay, FORMat, and SYSTem
15.
Section 6
(Buffer) for details.
Section
Section 12
(Common Commands) for
17 = MENU key
18 = COMM key
19 = DISP key
20 = TRIG key
21 = HALT key
22 = DIGITS key
23 = RATE key
24 = Cursor left arrow key
25 = -------
26 = SAVE key
27 = SETUP key
28 = STORE key
29 = RECALL key
30 = LIMIT key
31 = AZERO key
32 = EXIT key
13-9
15. A setup