Keithley 6485 Instruction Manual page 102

Picoammeter
Hide thumbs Also See for 6485:
Table of Contents

Advertisement

7-8
Triggering
Table 7-2 (continued)
Trigger model menu structure
- - DELAY
- - - MAN
- - - AUTO
- - TRIG-OUT
- - - LINE
- - - - TLINK
- - - EVENTS
- - - - VMC
- - TRIG-IN
- - - SOURCE
- - - - IMM
- - - - TLINK
- - - - -TLINK
- - - - -EVENT
- ARM
- - ARM-IN
- - - IMM
- - - TLINK
- - - - TLINK
- - - - EVENT
- - - MAN
- - - TIMER
- - - GPIB
- - ARM-OUT
- - - LINE
- - - - TLINK
- - - EVENT
- - - - T - L - DONE
- - COUNT
- - - INF
- - - FIN
NOTE
Menu
Path to ARM layer submenus
Input trigger and output triggers cannot share the same trigger link line.
Defaults set line 1 as the input and line 2 as the output.
Model 6485 Picoammeter Instruction Manual
Description
Set Trigger delay
Specify trigger delay 0–999.9998sec.
Enable auto delay
Configure output triggers
Select the output trigger link line
Set trigger link value: 1–6
Select VMC (voltmeter complete) output trigger
Enable/disable VMC (on or off)
Path to control source
Set the TRIG-IN control source
Set control source to IMMediate
Set control source to TLINK
Set TLINK value: 1–6
Enable (set to ONCE) or disable (set to NEVER)
Path to ARM-IN control source
Set control source to IMMediate
Set control source to TLINK
Set TLINK value: 1–6
Enable (set to ONCE) or disable (set to NEVER)
Set control source to MANual
Set control source to TIMER. Set timer from
0.001S to 27H, 46M, and 39.992S
Set control source to GPIB
Path to ARM-OUT trigger
Set the output trigger link line
Set trigger link value: 1–6
Set Trigger Layer Done event
Enable/disable Trigger Layer Done (on / off)
Set ARM measure count
Specify an INFinite measure count
Specify a FINite measure count: 1–2500

Advertisement