Connecting the Power Supply

Connecting the Power Supply

7 Install the main cover assembly. Confirm the cover stud guide pins are aligned properly with
the main chassis assembly.
8 Close the front access door.
9 Connect MS Foreline hose to the pump. Refer to pump documentation or more information.
10 If the Quiet Cover MS is not currently in its operating location, unlock the locking casters,
carefully push the Quiet Cover MS to its operating location, and lock the front locking casters.
Do not attempt to lift the Quiet Cover MS with the rough pump installed. Lifting the
WAR N IN G
Quiet Cover MS with the rough pump installed may result in personal injury.

Connecting the Power Supply

Connect the power supply to the back of Quiet Cover MS by twisting the ring on the circular
connector of the power cord
outlet.
Figure 9.
Quiet Cover MS User Manual
(Figure
9). Plug the pronged end of the power cable into the wall
Connecting the power supply
Connect
power
18