Replacing The Main Power Fuses - Agilent Technologies 1290 Infinity User Manual

Binary pump
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

9
Maintenance

Replacing the Main Power Fuses

Replacing the Main Power Fuses

When
If the main power LED is off while the main power button is pressed.
Tools required
Description
Flat head screwdriver
Parts required
#
1
Preparations
Switch off the instrument and unplug the main power cable.
Fire hazard
WA R N I N G
Using wrong fuses can result in fire hazard.
There are more fuses inside the instrument.
N O T E
If replacing the main power fuse does not resolve the issue, please contact your Agilent
service representative.
186
p/n
2110-1004
For continued protection against fire hazard, replace line fuses only with the same
type and ratings.
Only use the fuses specified for this instrument. The use of other fuses or materials
is prohibited.
Description
Fuse 10 A t
Agilent 1290 Infinity Binary Pump User Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents