Chapter 10 Monitoring SpO

10.1 Overview

SpO
is based on the absorption of pulse blood oxygen to red and infrared light by means of
2
finger sensor and SpO
determine the oxygen saturation of hemoglobin in the arterial blood. If, for example, 97% of the
hemoglobin molecules in the red blood cells of the arterial blood combine with oxygen, then the
blood has a SpO
oxygen saturation of 97%. The SpO
2
The SpO
numeric shows the percentage of hemoglobin molecules which have combined with
2
oxygen molecules to form oxyhemoglobin. The SpO
rate signal and a plethysmogram wave.
10.2 SpO
Safety Information
2
1
If the SpO
sensor can not work properly, please reconnect the sensor or change a
2
new one.
2
Do not use the sterile supplied SpO
damaged and return them to the vendor.
3
Prolonged and continuous monitoring may increase the risk of unexpected change of
dermal condition such as abnormal sensitivity, rubescence, vesicle, repressive
putrescence, and so on. It is especially important to check the sensor placement of
neonate and patient of poor perfusion or immature dermogram by light collimation and
proper attaching strictly according to changes of the skin. More frequent examinations
may be required for different patients.
4
Tissue damage may be caused by incorrect application or prolonged measurement
duration using the sensor (more than 4 hours). Inspect the sensor periodically
according to the sensor user manual.
5
Neonate SpO
2
time.
6
Use only EDAN permitted sensors and extension cables with the moniter. Other
sensors or extension cables may cause improper monitor performance and/or minor
personal injury.
NOTE:
1
Make sure the nail covers the light window. The wire should be on the backside of the
hand.
2
SpO
waveform is not proportional to the pulse volume.
2
3
Avoid placing the sensor on extremities with an arterial catheter, or intravascular
venous infusion line.
measuring unit. SpO
2
sensor can only be used when required, no more than 20 min at a
2
Plethysmogram measurement is employed to
2
numeric on the monitor will read 97%.
2
/PLETH parameter can also provide a pulse
2
WARNING
sensors if the packaging or the sensor is
2
- 70 -
Monitoring SpO
2