Recorder; Review; Ecg; 3, 5-Lead Monitoring - EDAN iM80 User Manual

Patient monitor
Hide thumbs Also See for iM80:
Table of Contents

Advertisement

Patient Monitor User Manual
A.6 Recorder
Record Width
Paper Speed
Trace
Recording types
A.7 Review
Trend Review
Short
Long

Review

A.8 ECG
A.8.1 3, 5-lead Monitoring
Lead Mode
Waveform
Lead naming style
Display Sensitivity
Bandwidth (-3dB)
CMRR
(Common
Rejection Ratio)
48 mm
25 mm/s, 50 mm/s
Up to 3
8 second real-time recording
Auto 8 second recording
Parameter alarm recording
Trend recording
Titration table recording
1 hrs, 1 Second Resolution
120 hrs, 1 min. Resolution
1200 sets NIBP measurement data
50 sets 12-lead ECG diagnosis results
3-Lead: I, II, III
5-Lead: I, II, III, aVR, aVL, aVF, V
3-Lead: 1-channel waveform
5-Lead: 2-channel waveform, max. seven waveforms;
AHA, IEC
1.25mm/mV (×0.125), 2.5mm/mV (×0.25), 5mm/mV
(×0.5), 10mm/mV (×1), 20mm/mV (×2), AUTO gain
Diagnosis: 0.05Hz to 150Hz
Monitor: 0.5Hz to 40Hz
Surgery: 1Hz to 20Hz
Diagnosis: >95dB (the Notch filter is off)
Mode
Monitor: >105dB (the Notch filter is on)
Surgery: >105dB (the Notch filter is on)
- 145 -
Product Specification