Chapter 26 Maintenance; Inspecting; Maintenance Task And Test Schedule - EDAN iM80 User Manual

Patient monitor
Hide thumbs Also See for iM80:
Table of Contents

Advertisement

Patient Monitor User Manual

Chapter 26 Maintenance

1
Failure on the part of the responsible individual hospital or institution employing the
use of this equipment to implement a satisfactory maintenance schedule may cause
undue equipment failure and possible health hazards.
2
If you discover a problem with any of the equipment, contact your service personnel,
or your authorized supplier.

26.1 Inspecting

The overall check of the monitor, including the safety check, should be performed only by
qualified personnel every 24 months, and each time after fix up.
The following items should be checked:
 If the environment condition and power supply meet requirement.
 If the power supply cord has damage and insulativity meets requirement.
 If the device and accessories have damage.
 Specified accessories.
 If the alarm system can work properly.
 If the recorder can work properly and the paper meets the requirement.
 Battery performance
 If all monitoring functions are in good conditions.
 If the grounding resistance and leakage current meet requirement.
If any damage or abnormality is found, please don't use the monitor and contact local Customer
Service Center.

26.2 Maintenance Task and Test Schedule

The following tasks are for EDAN-qualified service professionals only. Contact an
EDAN-qualified service provider if your monitor needs a safety or performance test. Clean and
disinfect equipment to decontaminate it before testing or maintaining it.
Maintenance and Test Schedule
Safety checks. Selected tests on
the basis of IEC60601-1
Check ECG synchronization of the
monitor and defibrillator
WARNING
Frequency
At least once every two years, or as needed, after any
repairs where the power supply is removed or replaced,
or if the monitor has been dropped.
At least once every two years, or as needed.
- 134 -
Maintenance

Advertisement

Table of Contents
loading

This manual is also suitable for:

Im50