EDAN iM80 User Manual page 44

Patient monitor
Hide thumbs Also See for iM80:
Table of Contents

Advertisement

Patient Monitor User Manual
NIBP Signal Sturated
NIBP
Init
Pressure
High
NIBP Invalid Reset
NIBP Time Out
NIBP Tube Leak
NIBP Cuff Type Error
Air Pressure Error
NIBP Self Test Error
Signal amplitude too strong
The initial pressure is too
high during measuring
The hardware pressure is
too high
Measuring
time
exceeded the specified time.
NIBP cuff or pump has a
leakage.
The cuff type used isn't
consistent with the patient
type.
Environment
atmospheric
pressure
abnormal
system pressure abnormal.
Sensor or other hardware
errors.
- 33 -
Low
High
Low
has
Low
Low
Low
or
Low
High
Alarm Information
Stop the patient from
moving.
Measure
again,
if
failure persists, stop
measuring
function
of NIBP module and
notify
biomedical
engineer
or
manufacturer's
service staff.
Measure
again,
if
failure persists, stop
using
measuring
function
of
NIBP
module and notify
biomedical engineer
or
manufacturer's
service staff.
Measure again or use
other
measuring
method.
Check the NIBP cuff
and
pump
for
leakages.
Confirm the patient
type and change the
cuff.
Check whether the
airway is occluded or
pressure
sensor
works properly in
pressure meter mode.
If the problem still
exists, contact your
service personnel.
Contact your service
personnel.