Patient Monitor User Manual
NOTE:
1
Always connect the airway adapter to the sensor before inserting the airway adapter
into the breathing circuit. In reverse, always remove the airway adapter from the
breathing circuit before removing the sensor.
2
Always disconnect the cannula, airway adapter or sample line from the sensor when
the sensor is not in use.
16.3.2.2 Removing Exhaust Gases from the System
Anesthetics: When using the sidestream CO
receiving or have recently received anesthetics, connect the outlet to a scavenging
system, to avoid exposing medical staff to anesthetics.
Use an exhaust tube to remove the sample gas to a scavenging system. Attach it to the sidestream
sensor at the outlet connector.
16.3.3 C5 CO
2
NOTE:
You must perform a zero calibration as described in this procedure each time you use a
new airway adapter.
16.3.3.1 Measurement Steps
1
Attach the sensor connector to the CO
2
Wait two minutes, allowing the sensor to reach its operating temperature and a stable
thermal condition.
3
Choose the appropriate airway adapter and connect it to the sensor head. The airway adapter
clicks into place when seated correctly.
Module
C5 CO
WARNING
measurement on patients who are
2
module
2
connector on the monitor.
2
- 92 -
Monitoring CO