EDAN iM80 User Manual page 45

Patient monitor
Hide thumbs Also See for iM80:
Table of Contents

Advertisement

Patient Monitor User Manual
NIBP Pneumatic Leak
NIBP System Failure
NIBP Cuff Leak
NIBP Leak Test Error
NIBP Weak Signal
NIBP
Excessive
Motion
NIBP Range Exceeded
NIBP Air Leak
NIBP Pressure Low
NIBP cuff or pump has a
leakage.
Hardware abnormal
Cuff, pump or airway has a
leakage.
Hardware abnormal
Cuff is too loose or patient
pulse is too weak.
Due to arm motion, signal
noise is too large or pulse
rate is not regular.
Maybe the patient blood
pressure value is beyond the
measurement range.
The cuff pressure can't reach
the set value within 60sec
and 20sec in Neo mode.
The module isn't able to
detect
the
SYSTOLIC.
Apply pressure again.
- 34 -
Check the NIBP cuff
Low
and
leakages.
Contact your service
High
personnel.
NIBP
properly connected,
Low
or there is a leak in
the airway.
Check whether the
airway is occluded or
pressure
works properly in
High
pressure meter mode.
If the problem still
exists, contact your
service personnel.
Use other method to
Low
measure
pressure.
Make sure that the
patient
Low
monitoring
motionless.
Maybe the patient
blood pressure value
Low
is
measurement range.
Check
connections and the
Low
wrapped cuff to see
whether they are all
prepared well.
Inflate
retry thrice. Check
whether the patient
Low
has an over high
blood pressure or it
is
movement.
Alarm Information
pump
for
cuff
isn't
sensor
blood
under
is
beyond
the
the
again
and
interfered
by