Patient Monitor User Manual
Normal
Brady
Range of Ventricular Rhythm
Ventricular Tachycardia
Ventricular Rhythm
Ventricular Bradycardia
Startup time for Tachycardia
Ventricular Tachycardia
1 mV 206bpm
Ventricular Tachycardia
2 mV 195bpm
Response time of Heart
Rate Meter to Change in
HR
Tall T-wave Rejection
Accuracy of Heart Rate
Meter and Response to
Irregular Rhythm
16
different
arrhythmia
analyses
ADU: 41 bpm ~ 119 bpm
PED/NEO: 61 bpm ~159 bpm
ADU: 15 bpm ~ 40 bpm
PED/NEO: 15 bpm ~ 60 bpm
The interval of 5 consecutive ventricular wave is less than 600 ms
The interval of 5 consecutive ventricular wave ranges from 600
ms to 1000 ms
The interval of 5 consecutive ventricular wave is more than 1000
ms
Gain 1.0: 10 s
Gain 0.5: 10 s
Gain 2.0: 10 s
Gain 1.0: 10 s
Gain 0.5: 10 s
Gain 2.0: 10 s
HR range: 80 bpm ~ 120 bpm
Range : 7s ~ 8s, average is 7.5s
HR range: 80bpm ~ 40bpm
Range : 7s ~ 8s, average is 7.5s
Exceeds ANSI/AAMI EC13-2002 Sect. 4.1.2.1 C) minimum
recommended 1.2mV T-Wave amplitude
Complies with ANSI/AAMI EC13-2002 Sect.4.1.2.1 e)
The HR value displays after a stable period of 20s:
According with ANSI/AAMI EC13-2002 Sect.4.1.2.1 e)
Ventricular bigeminy: 80bpm±1bpm
Slow alternating ventricular bigeminy: 60bpm±1bpm
Rapid alternating ventricular bigeminy: 120bpm±1bpm
Bidirectional systoles: 91bpm±1bpm
Non-Paced Patient
ASYSTOLE
VFIB/VTAC
- 152 -
R on T
PVC
Product Specification
Paced Patient
ASYSTOLE
TACHY