Resp; Nibp; Nibp From Edan Module - EDAN iM80 User Manual

Patient monitor
Hide thumbs Also See for iM80:
Table of Contents

Advertisement

Patient Monitor User Manual
A.9 RESP
Method
Respiration excitation
waveform
Measuring sensitivity
Base impedance range
Differential input impendence
Waveform bandwidth
RR measuring range:
Adult
Neo/Ped
Resolution
Accuracy
Apnea Alarm Time Setup
A.10 NIBP
A.10.1 NIBP from EDAN Module
Method
Mode
Measuring interval in AUTO
Mode
Continuous
Measuring type
Measuring Range
Adult mode
Pediatric mode
Neonatal mode
Trans-thoracic impedance: R-F(RA-LL), R-L (RA-LA)
< 300 μA, sinusoid, 62.8 kHz (± 10%)
0.3 Ω (base impedance 200 to 4500 Ω)
200 to 2500 Ω (cable resistance = 0 K)
2200 to 4500 Ω (leads cables 1KΩ resistance)
> 2.5MΩ
0.2 to 2.5 Hz (-3 dB)
0 to 120 rpm
0 to 150 rpm
1 rpm
±2 rpm
10s, 15s, 20s, 25s, 30s, 35s, 40s
Oscillometric
Manual, Auto, Continuous
1/2/3/4/5/10/15/30/60/90/120/240/480 min
5min, interval is 5s
SYS, DIA, MAP
SYS: 40 mmHg ~ 270 mmHg
DIA: 10 mmHg ~ 215 mmHg
MAP: 20 mmHg ~ 235 mmHg
SYS: 40 mmHg ~ 200 mmHg
DIA: 10 mmHg ~ 150 mmHg
MAP: 20 mmHg ~ 165 mmHg
SYS: 40 mmHg ~ 135 mmHg
DIA: 10 mmHg ~ 100 mmHg
- 154 -
Product Specification

Advertisement

Table of Contents
loading

This manual is also suitable for:

Im50

Table of Contents