Patient Monitor User Manual
2. Select Menu > Common Function, then choose Demo Mode from the popup interface and
input password 3045.
After entering Demo Mode, the monitor will perform the followings:
 Storing no data of new patient.
 Pausing to give all types of alarm.
 Pausing to transmit patient data to CMS and other network devices.
 Pausing to store the currently recorded data, and clearing the memory used to store
recording and printing data.
 Real data: the parameter measurement value and real waveform displayed on the screen
are from the predefined analog data, not the truly monitoring patients' data.
 History data: the monitor will store the analog real-time data in Demo mode, including
trend data, patient information, alarm event, waveform and setting.
To exit Demo Mode, select Menu > Common Function > Demo Mode to exit.
This is for demonstration purposes only. You must not change into Demo Mode during
monitoring. In Demo Mode, all stored trend information is deleted from the monitor's
memory.

3.3.2 Standby Mode

Standby mode can be used when you want to temporarily interrupt monitoring. To enter standby
mode, please press the shortcut key
anything on the screen or press any key.

3.4 Changing Monitor Settings

3.4.1 Adjusting Screen Brightness

To change the screen brightness, please:
1. Press the Brightness key on the screen directly or.
2. Select Menu > Common Function > Brightness, and select the appropriate setting for the
screen brightness. 10 is the brightest, 1 is the least bright.
Your monitor may be configured with lower brightness in standby mode and also for transport to
conserve battery power.

3.4.2 Changing Date and Time

To change the date and time, please refer to Section Setting Date and Time.
WARNING
on the screen directly. To resume monitoring, select
- 19 -
Basic Operation

Advertisement