16.3 Monitoring Procedures

16.3.1 Zeroing the sensor

You must perform zeroing following the steps when using the new airway adapter.
1. Expose the sensor to room air and keep it away from all sources of CO
ventilator, the patient's breath and the operator's.
2. In the CO
2
3. In the CO
2
4. If the system briefly displays Zero In Progress, the process is successful. After the
zeroing calibration is finished, you can start CO
Breath detected or Zero required, zeroing has failed. Zero calibration must be
performed again.
16.3.2 LoFlo CO
NOTE:
You must perform a zero calibration as described in this procedure each time the ambient
temperature changes more than 10°C (for example during transport).
16.3.2.1 Measurement Steps
1
Plug the sensor cable into the monitor's CO
Setup menu, please set the Work Mode to Measure.
Setup menu, select Zero Calibration.
Module
2
LoFlo CO
Monitoring. If the system displays
2
module
2
input connector. Allow the sensor two minutes
2
- 90 -
Monitoring CO
measurement, or the
2
including the
2
2