C.o - EDAN iM80 User Manual

Patient monitor
Hide thumbs Also See for iM80:
Table of Contents

Advertisement

Patient Monitor User Manual
Enflurane
Isoflurane
Sevoflurane
Xenon
Helium
Desflurane
Barometric Pressure on EtCO
Quantitative effect
Ambient Barometric, Operational
0 – 40 mmHg: ± 1 mmHg additional error
41 – 70 mmHg: ± 2.5% additional error
71 – 100 mmHg: ± 4% additional error
101 – 150 mmHg: ± 5% additional error
*Additional worst case error when compensation for P
is correctly selected for the actual fractional gas constituents present.
A.16 C.O.
Method
Measuring range

C.O.

TB
TI
Resolution
C.O.
TB, TI
Accuracy
C.O.
5
41 – 70 mmHg: ± 2.5% additional error
5
71 – 100 mmHg: ± 4% additional error
5
101 – 150 mmHg: ± 5% additional error
80
*Additional worst case error when compensation
for P
50
correctly selected for the actual fractional gas
15
constituents present.
Desflurane:
The presence of desflurane in the exhaled breath at
concentrations greater than 5% will positively bias
Carbon Dioxide values by up to an additional 3
mmHg at 38mmHg.
Xenon:
The presence of Xenon in the exhaled breath will
negatively bias Carbon Dioxide values by up to an
additional 5 mmHg at 38mmHg.
Measurement Values:
2
Thermodilution Technique
0.1 L/min ~ 20L/min
23°C ~ 43°C(73.4 º F ~109.4 º F)
-1°C ~ 27°C(30.2 º F ~80.6 º F)
0.1L/min
+0.1°C (+0.1 º F)
For C.O. ≥ 4.0 L/min: ±5%
- 160 -
, O
, N
O, anesthetic agents, or helium is
B
2
2
, O
, N
O, anesthetic agents, or helium
B
2
2
Product Specification