EDAN iM80 User Manual page 176

Patient monitor
Hide thumbs Also See for iM80:
Table of Contents

Advertisement

Patient Monitor User Manual
CO
2
N
O
2
Agents
Apnea Alarm
Delay
Temperature(°C)
Relative
humidity
Atmospheric
pressure
Interfering gas and vapour effects
Gas or vapour
4)
N
O
2
4)
HAL
4)
ENF, ISO, SEV
4)
DES
4)
Xe(Xenon)
4)
He(Helium)
Metered dose
inhaler
4)
propellants
C
H
OH(Ethanol)
2
5
4)
C
H
OH
3
7
4)
(Isopropanol)
CH
COCH
3
3
4)
(Acetone)
4)
CH
(Methane)
4
CO(Carbon
5)
monoxide)
±(0.3 vol% + 4 % of reading)
±(2 vol% + 4 % of reading)
±(0.2 vol% + 10 % of reading)
20s ~ 60s
10 to 40
10 ~ 95% RH(non-condensing )
525 ~ 1200hPa
Gas level
CO
2
IRMA CO
_ 1&2)
60 vol%
_ 1)
4 vol%
5 vol%
+8% of
reading
15 vol%
+12% of
reading
80 vol%
-10% of
reading
50 vol%
-6% of
reading
Not for use with metered dose inhaler propellants
_ 1)
0.3 vol%
_ 1)
0.5 vol%
_ 1)
1 vol%
_ 1)
3 vol%
_ 1)
1 vol%
-20 ~ 75
5 ~ 100% RH(condensing )
500 ~ 1200hPa
IRMA AX+
2
_ 1&2)
_ 1)
_ 1)
3)
_ 1)
3)
3)
3)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
- 165 -
Product Specification
Agents
N
O
2
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)
_ 1)

Advertisement