This mode can only be activated in measuring function
«Measurement with Statistics».
2-Layer-Correction
4.5

Language

The language of the display texts can be selected.
4.6

Basic Setups

The following settings are possible:
- Unit
If you choose «mm» the bar diameter setting can be
selected in mm and the results will be indicated in the
metric system. If you chose «inch» the bar diameter
setting can be made in bar size numbers and the
results are indicated in imperial units. Please refer also
to «Bar Diameter» on page 6.
8
Settings
Warning!
By changing the unit from mm to inch or vice
versa existing measured and stored values are
falsified. In order to avoid confusing results the
user must clear the memory after changing the
units.
The mode «2-Layer-Correction» can only be used for
metric units.
- Probe
When starting up the device for the first time,
the code no. engraved on the probe must be
entered in the submenu «Probe».
- Audible locating aid
(Short) beep tone selectable in two frequencies or
Variotone can be selected.
The remaining settings are described under the respec-
tive functions.
4.7

Data Output

For description, please see under «Data Output»,
page 20.
4.8
Measure with Statistics
For description, please see under «Measure with
Statistics», page 9.