Measure with Grid

Model SCANLOG only
This function displays concrete covers over a large area
in grey scale or colour shades. The smallest concrete
cover measured in a grid field is displayed.
Settings
See also «Settings», page 6.
• Enter the bar diameter of the first layer.
• Enter the object number.
• Enter the bar spacing under the menu option
«Corrections»
→ «Neighb. Bar Corr.» if the correction is required.
The «2-Layer Correction» can not be used by this
function.
• Select the size of the grid field («MENU» → «Basic
setups» → «xy-grid»).
The grid should usually be greater than the division of
the first layer (e.g. +50 %) to ensure that at least one
rebar is within a grid field.
• Set the range in which the concrete covers are to be
displayed in grey scale («MENU» → «Basic setups»
→ «Display»).
Change the grey scale range to improve the contrast of
the displayed object (see Fig. 5.12).
© 2007 Proceq SA
Fig. 5.12 Adjusting grey scale range
The range can be adjusted as desired even after
measuring.
• Select the «Measure with Grid» function.
• Pressing the START key calls the measuring display:
Fig. 5.13 «Measure with Grid» measuring display
Measuring Process
17