Inspection And Connection Of Ancillary Equipment - Olympus BF-UC180F Instructions Manual

Bf type evis exera ii ultrasonic bronchofibervideoscope
Hide thumbs Also See for BF-UC180F:
Table of Contents

Advertisement

Chapter 3 Preparation and Inspection
3.5
Inspection and connection of ancillary
equipment
Preparation and inspection of ancillary equipment
Connection of the endoscope and endoscopic equipment
38
This endoscope is not compatible with the video system
center CV-100/200/240/260. Use the video system center
CV-140/160/180.
1.
Prepare and inspect the light source, endoscopic ultrasound center, video
system center, monitor, suction pump, and EndoTherapy accessories as
described in their respective instruction manuals.
2.
Confirm that there are no scratches, cracks, excessive wear, or deformation
of the ultrasonic cable.
If the endoscope connector, videoscope cable and video
system center are not connected properly, the endoscopic
image may flicker or not be displayed. Continuous use of
such an endoscope may cause patient injury, bleeding, or
perforation.
Firmly connect the suction tube from the suction pump to the
suction connector on the suction valve. If the suction tube is
not attached properly, debris may drip from the tube. The
patient, operator, and/or equipment could be contaminated
and equipment malfunction can result.
For more information on combining the endoscope with the
video system center and the videoscope cable, refer to the
"System chart" in the Appendix.
1.
If any ancillary equipment is ON, turn it OFF.
2.
Insert the endoscope connector completely into the output socket of the light
source.
3.
Make sure that the inside of the videoscope cable connector is dry and free
of debris.
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F