2.3

Specifications

ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
8. Bending section
This section moves the distal end of the endoscope when the UP/DOWN
angulation control lever is operated.
9. Remote switches
The functions of remote switches 1 to 4 can be selected on the video system
center. When selecting the functions, refer to the instruction manual for the
video system center.
10. UP/DOWN angulation control lever
When this lever is turned in the "U" direction, the bending section moves UP.
When the lever is turned in the "D" direction, the bending section moves
DOWN.
Compared with the BF-UC160F-OL8, the channel inner
diameter of the BF-UC180F is increased from 2.0 mm to
2.2 mm. However, the compatible EndoTherapy accessories
of the BF-UC180F are same as that of the BF-UC160F-OL8.
This endoscope differs from a general videoscope. The
endoscopic image is transmitted to the control section by the
image guide and converted into an electrical signal with the
CCD in the control section.
Only a limited type of EndoTherapy accessories are
compatible with this endoscope compared with a general
ø 2.2 mm channel endoscope. This is because it has a
unique shaped instrument channel outlet. Refer to the
"System chart" in the Appendix to confirm the detail about the
compatible EndoTherapy accessories.
Chapter 2 Instrument Nomenclature and Specifications