Required Reprocessing Equipment - Olympus BF-UC180F Instructions Manual

Bf type evis exera ii ultrasonic bronchofibervideoscope
Hide thumbs Also See for BF-UC180F:
Table of Contents

Advertisement

7.1

Required reprocessing equipment

Preparation of the equipment
Channel cleaning
1
brush (BW-7B
) or
single use single-ended
cleaning brush
1
(BW-400B
)
Channel-opening cleaning brush
(MH-507) or
single use channel-opening cleaning
brush (MAJ-1339, 3 pcs
Water-resistant cap
(MH-553, 2 pcs)
• Detergent solution
3
• 500 cm
(500 ml) containers
• Clean water (see Section 6.2)
• Sterile water
• 70% ethyl or isopropyl alcohol
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Prior to cleaning and disinfection or sterilization, prepare the equipment shown in
Figure 7.1.
Use basins that are at least 40 cm by 40 cm (16" by 16") in
size and deep enough to allow the endoscope to be
completely immersed.
Channel cleaning brush
(BW-15B) or
single-use channel cleaning
brush (BW-201B, 3 pcs
1
)
Maintenance unit
(MU-1)
1
These products may not be available in some areas.
• Disinfectant solution
• Large non-metallic basin with a
tight fitting lid for detergent and
disinfectant solution
• Large basin for rinsing and
leakage testing
3
• 30 cm
(30 ml) syringe
• Soft brush
Figure 7.1
Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
Single use combination
cleaning brush
1
(BW-411B
)
1
)
Leakage tester
(MB-155)
Suction cleaning adapter
(MAJ-222)
Suction pump
(KV-4, KV-5, SSU-2)
• Personal protective equipment
• Clean, lint-free cloth
• Sterile, lint-free cloth
• Sterile, cotton swabs
• Small containers
83