3.5
Inspection and connection of ancillary
equipment
Inspection of ancillary equipment
Connection of the endoscope and ancillary equipment
EVIS EXERA TJF TYPE 160VR OPERATION MANUAL
Attach the water container to the specified receptacle on the
trolley or the light source. If the water container is attached
anywhere else, water may drip from the water container's
water supply tube, and equipment malfunction can result.
Take care not to spill water from the water container's
connection adapter when detaching the connection adapter
from the endoscope. Spilled water could splash on the
equipment, and it may cause equipment malfunction.
Prepare and inspect the light source, video system center, video monitor, water
container, suction pump and endo-therapy accessories as described in their
respective instruction manuals.
Firmly connect the suction tube from the suction pump to the
suction connector on the endoscope connector. If the suction
tube is not attached properly, debris may drip from the tube
and can present an infection-control risk, damage and/or
reduce the performance.
Chapter 3 Preparation and Inspection
37