Advertisement

Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
7.5

Manual cleaning

Reusable part that is normally reprocessed with the
endoscope
Equipment needed
102
After completing the leakage test, perform manual cleaning according to the
procedures described below.
In case of excessive bleeding and/or delayed reprocessing, perform "Presoak for
excessive bleeding and/or delayed reprocessing" on page 113 and follow the
procedure described below.
Water-resistant cap (2 pcs)
Chain for water-resistance caps
Prepare the following equipment and wear appropriate personal protective
equipment.
• Personal protective equipment
• Soft brush
3
• 30 cm
(30 ml) syringe
• Clean, lint-free cloths
• Large basins
• Detergent solution
• Clean water (see Section 6.2)
• Suction cleaning adapter
• Channel cleaning brush or
single-use channel cleaning brush
• Channel-opening cleaning brush or
single use channel-opening cleaning
brush
• Channel cleaning brush
• Suction pump
• Single use combination cleaning brush
• Single use single-ended cleaning brush
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
(MH-553)
(MAJ-1723)
(MAJ-222)
(BW-15B)
(BW-201B)
(MH-507)
(MAJ-1339)
(BW-7B)
(BW-411B)
(BW-400B)