7.10 Care of the ultrasonic cable (MAJ-1597,
MAJ-1722)
Equipment needed
Care
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
• Clean, lint-free cloth
• Neutral detergent
• 70% ethyl or isopropyl alcohol
1.
If the cable is soiled with blood or other potentially infectious materials, wipe
off all gross debris using a soft, lint-free cloth moistened with neutral
detergent.
2.
To remove dust, dirt and other non-patient debris, wipe the cable's surface
with a soft, lint-free cloth moistened with 70% ethyl or isopropyl alcohol.
3.
Make sure that the cable is completely dry.
Never apply water to the ultrasonic cable. Equipment
damage can result.
The ultrasonic cable is not waterproof. Never immerse it in
disinfectant solution or any other fluids.
Alcohol is not a sterilant or high-level disinfectant.