Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
96
Figure 7.14
3.
Hold the pull-tab of the single use adapter biopsy valve (MAJ-1414) and pull
it as shown in Figure 7.15 below to break it and discard it.
Adapter biopsy
valve
Figure 7.15
4.
When using the single-use biopsy valve (MAJ-210), remove it as shown in
Figure 7.16 below and discard it.
When using the biopsy valve (MD-495), remove it and place it in a container
with detergent solution. Clean, disinfect or sterilize the items according to
the instructions given in Section 7.9, "Cleaning, disinfection, and sterilization
procedures for reusable parts and cleaning equipment".
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Suction
connector
Slit
Pull-tab