EVIS EXERA GASTROINTESTINAL VIDEOSCOPE
OLYMPUS GIF TYPE XP160
OLYMPUS GIF TYPE 160
OLYMPUS GIF TYPE Q160
OLYMPUS GIF TYPE Q160Z
OLYMPUS GIF TYPE 1TQ160
OLYMPUS GIF TYPE XTQ160
OLYMPUS CF TYPE Q160L/I
OLYMPUS CF TYPE Q160AL/I
OLYMPUS CF TYPE Q160ZL/I
OLYMPUS CF TYPE Q160DL/I
OLYMPUS PCF TYPE 160AL/I
OLYMPUS CF TYPE Q160S
Refer to the endoscope's companion manual, the "OPERATION MANUAL" whose cover lists the model of
your endoscope, for operation information.
USA: CAUTION: Federal law restricts this device to sale by or on the order of a physician.
EVIS EXERA COLONOVIDEOSCOPE
EVIS EXERA SIGMOIDOVIDEOSCOPE

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GIF XP160 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Olympus GIF XP160

  • Page 1 INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE XP160 OLYMPUS GIF TYPE 160 OLYMPUS GIF TYPE Q160 OLYMPUS GIF TYPE Q160Z OLYMPUS GIF TYPE 1TQ160 OLYMPUS GIF TYPE XTQ160 EVIS EXERA COLONOVIDEOSCOPE OLYMPUS CF TYPE Q160L/I OLYMPUS CF TYPE Q160AL/I...
  • Page 3: Table Of Contents

    Contents Contents Symbols..................Chapter 1 General Policy ............Instructions..................Signal words ................... Precautions ..................Chapter 2 Compatible Reprocessing Methods and Chemical Agents ..............Compatibility summary..............Detergent solution................Disinfectant solution................ Rinsing water .................. ETO gas sterilization............... Steam sterilization (autoclaving) of accessories ......Chapter 3 Cleaning, Disinfection and Sterilization Procedures ..................
  • Page 4 Contents EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL...
  • Page 5: Symbols

    Symbols Symbols The meaning(s) of the symbol(s) shown on the back cover of this instruction manual are as follows: Manufacturer Authorized representative in the European Community EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL...
  • Page 6: Chapter 1 General Policy

    • If you have any questions or comments about any information in this manual, or if a problem that cannot be solved occurs while reprocessing, contact Olympus. • The medical literature reports incidents of patient cross contamination resulting from improper cleaning, disinfection or sterilization. It is...
  • Page 7: Signal Words

    Chapter 1 General Policy Signal words The following signal words are used throughout this manual: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
  • Page 8 • Olympus confirms validation of the endoscope reprocessors recommended by Olympus only. If using endoscope reprocessor that are not recommended by Olympus, the endoscope reprocessor manufacturers are responsible for validation of the endoscope reprocessor with the endoscope models listed in its intended use statement. If using an endoscope reprocessor, confirm that it is capable of reprocessing endoscope including all channels.
  • Page 9 For information on the durability against each method, please contact Olympus. If cleaning, disinfection and sterilization methods not stated in this instruction manual are performed, Olympus cannot guarantee the effectiveness, safety and durability of this instrument.
  • Page 10 Chapter 1 General Policy • When reprocessing an endoscope that has the flexibility adjustment mechanism, make sure that the insertion tube is set to the softest condition (indicated by the “” mark on the flexibility adjustment ring) before immersing the endoscope in cleaning/disinfectant solution.
  • Page 11: Chapter 2 Compatible Reprocessing Methods And Chemical Agents

    Methods and Chemical Agents Compatibility summary Olympus endoscopic equipment is compatible with several methods of reprocessing. Certain components and accessories, however, are not compatible with some methods, which can cause equipment damage. For appropriate reprocessing methods, refer to Table 2.1 below, the recommendations of your infection control committee and all national and local hospital guidelines and policies.
  • Page 12 Chapter 2 Compatible Reprocessing Methods and Chemical Agents Steam sterilization (autoclaving) ETO gas sterilization 2 – 3.5% glutaraldehyde 70% ethyl or isopropyl alcohol Detergent solution Ultrasonic cleaning Endoscope Water-resistant cap (MH-553) Water-resistant cap (MAJ-583) Water-resistant cap (MAJ-942) Forceps suction plug (T-plug) (MH-405) Channel cleaning brush (BW-17K, BW-20T)
  • Page 13: Detergent Solution

    Use a medical-grade, low-foaming, neutral pH detergent or enzymatic detergent and follow the manufacturer’s dilution and temperature recommendations. Contact Olympus for the names of specific brands that have been tested for compatibility with the endoscope. Do not reuse detergent solutions.
  • Page 14: Rinsing Water

    Chapter 2 Compatible Reprocessing Methods and Chemical Agents Rinsing water Once removed from disinfectant solution, the instrument must be thoroughly rinsed with sterile water to remove any disinfectant residue. If sterile water is not available, clean potable tap water or water which has been processed (e.g. filtered) to improve its microbiological quality may be used.
  • Page 15 Chapter 2 Compatible Reprocessing Methods and Chemical Agents • Before sterilization, the instrument must be thoroughly cleaned and dried. Residual moisture inhibits sterilization. • The results of sterilization depend on various factors such as how the sterilized instrument was packed or the positioning, method of placing and loading of the instrument in the sterilization device.
  • Page 16: Steam Sterilization (Autoclaving) Of Accessories

    Chapter 2 Compatible Reprocessing Methods and Chemical Agents  Gas mixture For the USA 12% ETO/88% CFC For countries other than the USA 20% ETO/80% CO Steam sterilization (autoclaving) of accessories The accessories listed as compatible with steam sterilization (autoclaving) in Table 2.1 can be sterilized by steam within the parameters given in Table 2.3.
  • Page 17 Chapter 2 Compatible Reprocessing Methods and Chemical Agents • Allow the packages to dry within the autoclave, using the autoclave’s drying cycle (if available) or by opening the door of the autoclave and allowing the packages to air-dry. Handling a wet package can compromise its sterility. Process Parameters Prevacuum...
  • Page 18: Required Reprocessing Equipment

    Chapter 3 Cleaning, Disinfection and Sterilization Procedures Chapter 3 Cleaning, Disinfection and Sterilization Procedures ALL channels of the endoscope, including auxiliary water channel where fitted, MUST be cleaned and high-level disinfected or sterilized during EVERY reprocessing cycle, even if the channels were not used during the previous patient procedure.
  • Page 19 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Water-resistant cap (MH-553) Water-resistant cap Water-resistant cap (MAJ-942 (MAJ-583 for CF-Q160DL/I only) for CF-Q160ZL/I only) Channel-opening Channel cleaning brush cleaning brush (BW-20T) (MH-507) (BW-17K for GIF-XTQ160 only) Biopsy valve (MB-358) Channel plug (MH-944) Auxiliary water tube (MAJ-855 Suction cleaning adapter...
  • Page 20 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Reprocessing equipment parts and functions For inspection of other equipment than that mentioned below, refer to the instruction manual for the equipment being used.  Water-resistant cap (MH-553) The water-resistant cap is attached to the electrical connector on the endoscope to protect the connector from water penetration during reprocessing.
  • Page 21 Chapter 3 Cleaning, Disinfection and Sterilization Procedures  Water-resistant cap (MAJ-942 for CF-Q160DL/I only) The water-resistant cap (MAJ-942) is attached to the UPD scope connector on the endoscope to protect the connector from water penetration during reprocessing. The water-resistant cap is connected to the endoscope by a chain (see Figure 3.4).
  • Page 22 Chapter 3 Cleaning, Disinfection and Sterilization Procedures  Injection tube (MH-946) The injection tube is used to inject detergent solution, disinfectant solution, water and alcohol into the air/water and suction channels and to flush air through the channels to expel fluids (see Figure 3.6). Suction channel tube Notice card...
  • Page 23 Chapter 3 Cleaning, Disinfection and Sterilization Procedures  Channel cleaning brush (BW-17K, BW-20T) The channel cleaning brush is used to brush the inside of the instrument/suction channels and the holes in the air/water and suction valves (see Figure 3.8). Metal tip Figure 3.8 ...
  • Page 24 Chapter 3 Cleaning, Disinfection and Sterilization Procedures  Suction cleaning adapter (MH-856) The suction cleaning adapter is used to aspirate reprocessing fluids from the distal end of the endoscope through the instrument channel (see Figure 3.10). Weighted end Figure 3.10 ...
  • Page 25 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Inspection of the equipment For inspection of other equipment than that mentioned below, refer to the instruction manual for the equipment being used. The cleaning brushes are consumable items. Should the slightest irregularity be suspected, use a spare instead. Using a defective brush may cause equipment damage.
  • Page 26 Chapter 3 Cleaning, Disinfection and Sterilization Procedures  Inspection of the channel plug Confirm that the cylinder plug and biopsy valve cap are free from cracks, scratches, flaws and debris (see Figure 3.5).  Inspection of the injection tube Confirm that all components of the injection tube are free from cracks, scratches, flaws and debris (see Figure 3.6).
  • Page 27 Chapter 3 Cleaning, Disinfection and Sterilization Procedures  Inspection of the channel-opening cleaning brush Check for loose or missing bristles (see Figure 3.9). Check for debris on the shaft and/or in the bristles of the brush.  Inspection of the suction cleaning adapter Check for cracks, scratches, flaws, debris and other damage (see Figure 3.10).
  • Page 28: Cleaning, Disinfection And Sterilization Procedures

    Chapter 3 Cleaning, Disinfection and Sterilization Procedures Cleaning, disinfection and sterilization procedures Endoscope reprocessing summary chart Withdrawal of the endoscope Precleaning (Section 3.3) Leakage testing (Section 3.4) Manual cleaning (Section 3.5) High-level disinfection (Section 3.6) ETO gas sterilization (Section 3.8) Rinsing after high-level disinfection (Section 3.7) Storage and Disposal (Chapter 5)
  • Page 29: Precleaning

    Chapter 3 Cleaning, Disinfection and Sterilization Procedures ALL channels of the endoscope, including auxiliary water channel where fitted, MUST be cleaned and high-level disinfected or sterilized during EVERY reprocessing cycle, even if the channels were not used during the previous patient procedure.
  • Page 30 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Wipe down the insertion tube Handle the insertion tube carefully. Tightly gripping or sharply bending the insertion tube or bending section can stretch or severely damage the insertion tube and the covering of the bending section.
  • Page 31 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Flush water and air into the air/water channel Do not use the AW channel cleaning adapter for patient examinations. It will cause continuous insufflation and could result in patient injury. • To prevent clogging of the air/water nozzle, always use the AW channel cleaning adapter to clean the air/water channel after each use.
  • Page 32 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Attach the AW channel cleaning adapter to the air/water cylinder of the endoscope (see Figure 3.13). AW channel cleaning adapter Figure 3.13 Switch the airflow regulator to maximum output (“HIGH” or “3”). Immerse the distal end of the insertion tube in the water (see Figure 3.14). Depress the AW channel cleaning adapter to feed water through the channels for 30 seconds.
  • Page 33 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Flush detergent solution and air into the auxiliary water channel (for endoscopes with auxiliary water feeding only) • Do not cap the auxiliary water inlet during reprocessing. • If the auxiliary water tube is not connected to the auxiliary water inlet, uncap the auxiliary water inlet and connect the auxiliary water tube (see Figure 3.15).
  • Page 34 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Disconnect the endoscope, reusable parts and cleaning equipment Disconnect the AW channel cleaning adapter, suction valve and biopsy valve from the endoscope and place them in a container of detergent solution. Clean, disinfect and sterilize the items according to the instructions given in Section 3.9, “Cleaning, disinfection and sterilization procedures for reusable parts and reprocessing equipment”.
  • Page 35: Leakage Testing

    If the exterior of the electrical connector is scratched, the connector may no longer be waterproof and the seal inside the water-resistant cap may be scratched. If the electrical connector is scratched, send it immediately to Olympus for repairs. •...
  • Page 36 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Align the OER/ETD/EW character (see Figure 3.16 a), the EW character or KC/TD character (see Figure 3.16 b) on the water-resistant cap with mark 2 on the electrical connector housing. Align the pin on the electrical connector with the groove on the water-resistant cap.
  • Page 37 • If the seal ring on the UPD scope connector is scratched, the UPD scope connector may no longer be waterproof. Send the endoscope immediately to Olympus for repairs. • Never immerse the water-resistant cap unless it is attached to the endoscope. Water remaining inside the water-resistant cap can be transferred to and damage the UPD scope connector.
  • Page 38 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Zoom connector or UPD scope connector Seal ring Figure 3.17 Zoom connector or UPD scope connector No space Water-resistant cap Assembled Figure 3.18 EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL...
  • Page 39 This means that water will be able to penetrate the inside of the endoscope. If you locate a leak, remove the endoscope from the water and contact Olympus. • Never connect or disconnect the water-resistant cap(s) or the leakage tester’s connector cap while immersed.
  • Page 40 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Connect the leakage tester’s connector cap to the venting connector of the water-resistant cap (MH-553) (see Figures 3.19 or 3.20). Leakage tester Maintenance unit connector or light source Water-resistant cap (MH-553) Venting connector Leakage tester’s connector cap Figure 3.19...
  • Page 41: Manual Cleaning

    Chapter 3 Cleaning, Disinfection and Sterilization Procedures Wait 30 seconds, or until the covering of the bending section contracts to its pre-expansion size. Disconnect the leakage tester’s connector cap from the venting connector. Thoroughly dry the leakage tester. Manual cleaning After completing the leakage test, perform manual cleaning according to the procedures described below.
  • Page 42 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Cleaning the external surface Fill a basin with water and detergent solution at the temperature and concentration recommended by the detergent manufacturer. Use a basin which is at least 40 cm by 40 cm (16” by 16”) in size and deep enough to allow the endoscope to be completely immersed.
  • Page 43 Depending on the location, the missing part may not be recoverable by passing a new brush or other endo-therapy accessory through the channel. In this case, contact Olympus. Withdraw the channel cleaning brush from the suction channel gently, ensuring its shaft does not rub against the...
  • Page 44 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Suction cylinder Instrument channel port Suction connector Figure 3.22 Use the channel cleaning brush (BW-17K) to brush the instrument channel of the GIF-XTQ160. If the BW-17K is not used on this endoscope, effective cleaning of the instrument channel will not be possible.
  • Page 45 Chapter 3 Cleaning, Disinfection and Sterilization Procedures  Brushing the instrument/suction channel in the insertion tube (location A) (for GIF-XTQ160 only) Use the channel cleaning brush (BW-17K) to brush the instrument channel of the GIF-XTQ160. If the BW-17K is not used on this endoscope, effective cleaning of the instrument channel will not be possible.
  • Page 46 Chapter 3 Cleaning, Disinfection and Sterilization Procedures  Brushing the suction cylinder (location C) When inserting the channel-opening cleaning brush into the suction cylinder, do not forcibly insert the brush beyond the middle of the brush section. Otherwise, the brush may become stuck in the suction cylinder.
  • Page 47 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Reprocess the suction cleaning adapter as described in Section 3.9, “Cleaning, disinfection and sterilization procedures for reusable parts and reprocessing equipment”. Suction cleaning adapter Weighted end Figure 3.23 Instrument Weighted end channel port Suction channel Suction cleaning adapter Insertion tube...
  • Page 48 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Flushing detergent solution into the air/water channels Attach the channel plug’s biopsy valve cap to the instrument channel port (see Figure 3.25). Lower the channel plug’s cylinder plug onto the air/water and suction cylinders and slide the plug until it stops (see Figure 3.25).
  • Page 49 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Attach the 30 cm (30 ml) syringe to the injection tube’s air/water channel port (see Figure 3.27). Inject 90 cm (90 ml) of detergent solution into the air/water channel. Disconnect the channel plug and injection tube from the endoscope, and leave all items immersed.
  • Page 50 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Flushing detergent solution into the auxiliary water channel (for endoscopes with auxiliary water feeding only) Using the auxiliary water tube and 30 cm (30 ml) syringe, inject detergent solution into the auxiliary water channel. Attach the auxiliary water tube to the auxiliary water inlet (see Figures 3.28 and 3.29).
  • Page 51 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Soaking the endoscope and all equipment in detergent solution Using a clean, lint-free cloth, wipe all debris from the endoscope’s external surfaces while the endoscope is immersed in the detergent solution. Cover the basin with a tight fitting lid to minimize the release of detergent vapors.
  • Page 52 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Detach the channel plug and injection tube from the endoscope. Detach the auxiliary water tube from the endoscope (for endoscopes with auxiliary water feeding only). Channel plug Biopsy valve cap Suction cylinder Suction channel Suction channel Injection tube Suction connector...
  • Page 53 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Presoak for excessive bleeding and/or delayed reprocessing Follow the steps below only in case of excessive bleeding and/or delayed reprocessing; unnecessary immersions should be avoided. Consecutive extended immersions may damage the endoscope. Preclean and perform a leakage test on the endoscope as described in Section 3.3, “Precleaning”...
  • Page 54: High-Level Disinfection

    Chapter 3 Cleaning, Disinfection and Sterilization Procedures High-level disinfection After manual cleaning, disinfect the endoscope according to the procedures described below. Equipment needed Prepare the following equipment: • Personal protective equipment • Clean, lint-free cloths • Large basin • Disinfectant solution •...
  • Page 55 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Flushing disinfectant solution into all channels Immerse the endoscope and all equipment in the disinfectant solution. Using the 30 cm (30 ml) syringe, flush at least 90 cm (90 ml) of disinfectant solution through the air/water and suction channels respectively. Confirm that no bubbles exit the distal end of the endoscope.
  • Page 56: Rinsing After High-Level Disinfection

    Chapter 3 Cleaning, Disinfection and Sterilization Procedures Attach the 30 cm (30 ml) syringe to the luer port of the auxiliary water tube and flush the auxiliary water channel with 90 cm (90 ml) of air (for endoscopes with auxiliary water feeding only). Remove the endoscope and all equipment from the disinfectant solution.
  • Page 57 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Equipment needed Prepare the following equipment: • Personal protective equipment • Sterile, lint-free cloths • Large basin • Sterile water (for sterile water rinse) • Suction pump (with sterile suction tube) • 30 cm (30 ml) syringe •...
  • Page 58 Chapter 3 Cleaning, Disinfection and Sterilization Procedures  Sterile water rinse Fill a basin with sterile water. Use a basin which is at least 40 cm by 40 cm (16” by 16”) in size and deep enough to allow the endoscope to be completely immersed.
  • Page 59 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Flushing the channels with 70% ethyl or isopropyl alcohol after rinsing them with sterile water facilitates drying inside the channels.  Non-sterile water rinse and alcohol flush Alcohol is flammable. Handle with care. Fill a small container with 70% ethyl or isopropyl alcohol.
  • Page 60: Sterilization

    Chapter 3 Cleaning, Disinfection and Sterilization Procedures Disconnect all equipment from the endoscope. Using sterile cotton swabs, dry the inside of the air/water cylinder, suction cylinder and instrument channel port. Sterilization ETO gas sterilization As an alternative to high-level disinfection, the endoscope can be sterilized by ethylene oxide (ETO) gas.
  • Page 61 Chapter 3 Cleaning, Disinfection and Sterilization Procedures CF-Q160ZL/I, CF-Q160DL/I Figure 3.32 Seal the instrument in a package appropriate for ETO gas sterilization according to your hospital’s protocol. Sterilize the package according to the recommended ETO gas exposure parameters described in Section 2.5, “ETO gas sterilization” and the sterilizer manufacturer’s instructions.
  • Page 62: Cleaning, Disinfection And Sterilization Procedures For Reusable Parts And Reprocessing Equipment

    Chapter 3 Cleaning, Disinfection and Sterilization Procedures Cleaning, disinfection and sterilization procedures for reusable parts and reprocessing equipment This section includes the cleaning, disinfection and sterilization procedures for the reusable parts and reprocessing equipment listed below. For all other parts, refer to their respective instruction manuals.
  • Page 63 Chapter 3 Cleaning, Disinfection and Sterilization Procedures  Reusable parts which are normally reprocessed separately from the endoscope • Air/water valve (MH-438) • Suction valve (MH-443) • Biopsy valve (MB-358) • Mouthpiece (MA-474, MB-142) • AW channel cleaning adapter (MH-948) •...
  • Page 64 Chapter 3 Cleaning, Disinfection and Sterilization Procedures To clean the auxiliary water tube, attach the 30 cm (30 ml) syringe to the luer port of the auxiliary water tube and flush the tube thoroughly with detergent solution (for endoscopes with auxiliary water feeding only). Soak all equipment for the amount of time and at the temperature recommended by the detergent manufacturer.
  • Page 65 Chapter 3 Cleaning, Disinfection and Sterilization Procedures High-level disinfection Fill a basin with disinfectant solution at the temperature and concentration recommended by the disinfectant manufacturer. Use a basin which is deep enough to allow all equipment to be completely immersed. Immerse all equipment in the disinfectant solution.
  • Page 66 Chapter 3 Cleaning, Disinfection and Sterilization Procedures  Sterile water rinse Fill a basin with sterile water. Use a basin which is deep enough to allow all equipment to be completely immersed. Remove the suction cleaning adapter from the disinfectant solution and attach the 30 cm (30 ml) syringe.
  • Page 67 Chapter 3 Cleaning, Disinfection and Sterilization Procedures  Non-sterile water rinse and alcohol flush Alcohol is flammable. Handle with care. Fill a small container with 70% ethyl or isopropyl alcohol. Immerse all equipment in non-sterile water following all procedures given in “Sterile water rinse”...
  • Page 68 Chapter 3 Cleaning, Disinfection and Sterilization Procedures Sterilization  ETO gas sterilization The cleaning brushes are not compatible with ETO gas sterilization. After cleaning and drying as described in “Manual cleaning” on page 59, follow the procedures given below. Before sterilization, confirm that the parts of the equipment are thoroughly cleaned and dried.
  • Page 69: Chapter 4 Cleaning And Disinfection Equipment

    Chapter 4 Cleaning and Disinfection Equipment Chapter 4 Cleaning and Disinfection Equipment The endoscope is compatible with some endoscope reprocessors recommended by Olympus. Refer to the instruction manual of endoscope reprocessor for details on operation. • Thoroughly clean the endoscope as described in Chapter 3, “Cleaning, Disinfection and Sterilization Procedures”...
  • Page 70 Chapter 4 Cleaning and Disinfection Equipment • Use connectors special to each model when cleaning and disinfecting the endoscope with the endoscope reprocessor. Otherwise, insufficient cleaning and disinfection or sterilization of the endoscope may pose an infection control risk to the patient and/or operators performing the next procedure with the endoscope.
  • Page 71: Chapter 5 Storage And Disposal

    Chapter 5 Storage and Disposal Chapter 5 Storage and Disposal Storage • The storage cabinet must be clean, dry, well ventilated and maintained at ambient temperature. Storing the endoscope in direct sunlight, at high temperatures, in high humidity or exposed to X-rays and ultraviolet-rays may damage the endoscope or present an infection control risk.
  • Page 72: Disposal

    Chapter 5 Storage and Disposal Turn the endoscope’s angulation locks to the “F ” direction. Hang the endoscope in the storage cabinet with the distal end hanging freely. Make sure that the insertion tube hangs vertically and as straight as possible.
  • Page 73 ©2003 OLYMPUS MEDICAL SYSTEMS CORP. All rights reserved. No part of this publication may be reproduced or distributed without the express written permission of OLYMPUS MEDICAL SYSTEMS CORP. OLYMPUS is a registered trademark of OLYMPUS CORPORATION. Trademarks, product names, logos, or trade names used in this document are generally registered trademarks or trademarks of each company.
  • Page 74 A8F, Ping An International Financial Center, No. 1-3, Xinyuan South Road, Chaoyang District, Beijing, 100027 P.R.C. Fax: (86)10-5976-1299 Telephone: (86)10-5819-9000 Olympus Tower 9F, 446, Bongeunsa-ro, Gangnam-gu, Seoul, Korea 135-509 Fax: (02)6255-3494 Telephone: (02)6255-3210 491B, River Valley Road #12-01/04, Valley Point Office Tower, Singapore 248373...