3.6
Inspection of the endoscopic system
Inspection of the endoscopic image
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
SSD-5
Ultrasonic connector
ProSound F75
PROBE 2
PROBE 4
1
This product may not be available in some areas.
Figure 3.24
Do not stare directly at the distal end of the endoscope while
the examination light is ON. Otherwise, eye injury may result.
1.
Turn the video system center, light source, and monitor ON and inspect the
endoscopic image as described in their respective instruction manuals.
2.
Confirm that light is output from the endoscope's distal end.
3.
While observing the palm of your hand, confirm that the endoscopic image
is free from noise, blur, fog, or other irregularities.
Lock handle
1
Chapter 3 Preparation and Inspection
PROBE 1
PROBE 3
Lock handle
Ultrasonic connector
45